An unbiased approach to correct X-ray response non-uniformity in microstrip detectors has been developed based on the statistical estimation that the scattering intensity at a fixed angle from an object is expected to be constant within the Poisson noise. Raw scattering data of SiO glass measured by a microstrip detector module was found to show an accuracy of 12σ at an intensity of 10 photons, where σ is the standard deviation according to the Poisson noise. The conventional flat-field calibration has failed in correcting the data, whereas the alternative approach used in this article successfully improved the accuracy from 12σ to 2σ. This approach was applied to total-scattering data measured by a gapless 15-modular detector system. The quality of the data is evaluated in terms of the Bragg reflections of Si powder, the diffuse scattering of SiO glass, and the atomic pair distribution function of TiO nanoparticles and Ni powder.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6510202PMC
http://dx.doi.org/10.1107/S1600577519002145DOI Listing

Publication Analysis

Top Keywords

approach correct
8
correct x-ray
8
x-ray response
8
response non-uniformity
8
non-uniformity microstrip
8
microstrip detectors
8
poisson noise
8
sio glass
8
accuracy 12σ
8
statistical approach
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!