Purpose: Dynamic susceptibility contrast imaging requires high temporal sampling, which poses limits on achievable spatial coverage and resolution. Additionally, more encoding-intensive multi-echo acquisitions for quantitative imaging are desired to mitigate contrast leakage effects, which further limits spatial encoding. We present an accelerated sequence that provides whole-brain coverage at an improved spatio-temporal resolution, to allow for dynamic quantitative R and R * mapping during contrast-enhanced imaging.
Methods: A multi-echo spin and gradient-echo sequence was implemented with simultaneous multislice acquisition. Complementary k-space sampling between repetitions and joint virtual coil reconstruction were used along with a dynamic phase-matching technique to achieve high-quality reconstruction at 9-fold acceleration, which enabled 2 × 2 × 5 mm whole-brain imaging at TR of 1.5 to 1.7 seconds. The multi-echo images from this sequence were fit to achieve quantitative R and R * maps for each repetition, and subsequently used to find perfusion measures including cerebral blood flow and cerebral blood volume.
Results: Images reconstructed using joint virtual coil show improved image quality and g-factor compared with conventional reconstruction methods, resulting in improved quantitative maps with a 9-fold acceleration factor and whole-brain coverage during the dynamic perfusion acquisition.
Conclusion: The method presented shows the advantage of using a joint virtual coil-GRAPPA reconstruction to allow for high acceleration factors while maintaining reliable image quality for quantitative perfusion mapping, with the potential to improve tumor diagnostics and monitoring.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6692914 | PMC |
http://dx.doi.org/10.1002/mrm.27784 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!