The best intramedullary nail for the treatment of unstable AO/OTA 31.A2.1-3 fractures remains uncertain. A total of 237 patients (45 male, 192 female) were eligible for the assessment with an average age of 81.9 ± 10.5 years and a minimum follow-up of six months. We assessed the cut-out rate, the TAD and calTAD (Tip Apex distance) of three different implants. An overall cut-out rate of 2.5% ( = 6) was observed. The cut-out rate was 1.2% ( = 1) in the Proximal Femur Nail Antirotation (PFNA) group, 3.7% ( = 3) in the Gamma Nail group and 2.9% ( = 2) in the Gamma3 with a U-Blade (RC) lag-screw group. The TAD and calTAD differed between the cut-out and non-cut group-20.0 mm vs. 18.5 mm and 13.1 mm vs. 15.3 mm, respectively. A significantly higher TAD of 32.5 mm could be seen in the cut-out after PFNA ( < 0.0001). The only significant change in follow-up using Parker's ratio was observed in the PFNA group ( < 0.0001). The rate of patients requiring surgery after complications was 8.4% ( = 20) without any significant difference between the three groups. The PFNA blade showed significant migration within the femoral head, however the cut-out rate remained the smallest compared to Gamma3 with or without additional U-Blade (RC) lag screw.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6571935PMC
http://dx.doi.org/10.3390/jcm8050615DOI Listing

Publication Analysis

Top Keywords

cut-out rate
16
lag screw
12
ao/ota 31a21-3
8
gamma nail
8
u-blade lag
8
proximal femur
8
femur nail
8
nail antirotation
8
antirotation pfna
8
tad caltad
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!