A combined 32° full field of view refractive fundus camera and fixation target with a -20 to +10 diopter sphere correction range is described and demonstrated. The optical setup partially corrects the average longitudinal chromatic aberration and spherical aberration of the human eye, while providing a long eye relief to allow integration with reflective adaptive optics ophthalmoscopes, as a viewfinder. The fundus camera operates with 940 nm light, using a maximum 2.9 mm diameter imaging pupil at the eye. The fixation target uses a light projector capable of delivering red, green and/or blue spatially and temporally modulated stimuli to the retina. The design and performance of each sub-system are discussed, and retinal imaging at various wavelengths is demonstrated.