Objective: Develop sample acceptability rules by determining the relationship between free hemoglobin level (hemolysis) and potassium or ionized calcium in blood gas samples collected intraoperatively.
Design And Methods: Hemolysis was assessed visually or by H index for lithium heparin blood gas samples collected intraoperatively. During periods one and three this was done using two different rules for visual assessment of centrifuged lithium heparin plasma. During period two H index was measured for all visually hemolyzed samples on a Roche Cobas c501 analyzer to determine acceptability. Potassium and ionized calcium were measured in 75 lithium heparin whole blood samples on a Radiometer ABL90 to correlate H index and potassium or ionized calcium.
Results: During period one 35 of 5808 (0.6%) blood gas samples had visual hemolysis levels exceeding tolerance for reporting of potassium. By switching to measured H index using a laboratory-established threshold, during period 2 we estimate that 171 of 5396 (3.2%) blood gas samples exceeded the H index threshold for reporting of potassium. In 75 intraoperative blood gas samples with H index and whole blood potassium and ionized calcium measured; we observed no relationship between H index and potassium or ionized calcium. During period 3 we switched to visual assessment of hemolysis with a greater tolerance for hemolysis; with only 3 of 5345 (0.06%) samples exceeding the new visual hemolysis threshold.
Conclusion: For blood gas samples collected intraoperatively, there is no relationship between hemolysis and measured potassium or ionized calcium. The results suggest that only grossly hemolyzed intraoperative blood gas samples should be rejected for measurement of whole blood potassium and ionized calcium.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.clinbiochem.2019.05.002 | DOI Listing |
J Vet Emerg Crit Care (San Antonio)
January 2025
Department of Veterinary Surgical and Radiological Sciences, School of Veterinary Medicine, University of California, Davis, Davis, California, USA.
Background: Venous blood gas and acid-base analysis is often performed in cats and requires appropriate reference intervals (RIs) for interpretation. The currently available RIs were created from small numbers of cats, while current guidelines suggest that a minimum number of 40 samples is required for the creation of an accurate/appropriate RI.
Key Findings: In 41 healthy awake cats, pH, PCO, PO, sodium, potassium, ionized calcium, chloride, glucose, and plasma lactate concentrations were measured, while bicarbonate, standardized base excess, and anion gap were calculated.
J Phys Chem Lett
January 2025
Institute of Functional Nano & Soft Materials (FUNSOM), Jiangsu Key Laboratory for Carbon-Based Functional Materials & Devices, Joint International Research Laboratory of Carbon-Based Functional Materials and Devices, Soochow University, Suzhou 215123, P. R. China.
Alkali element doping has significant physical implications for two-dimensional materials, primarily by tuning the electronic structure and carrier concentration. It can enhance interface electronic interactions, providing opportunities for effective charge transfer at metal-organic interfaces. In this work, we investigated the effects of gradually increasing the level of K doping on the lattice structure and electronic properties of an organometallic coordinated Kagome lattice on a Ag(111) surface.
View Article and Find Full Text PDFCureus
November 2024
Infectious Diseases, Ziauddin University, Karachi, PAK.
Background Venous blood gas (VBG) investigation is a widespread option for arterial blood gas analysis because it is easier to draw and has a lower risk of complications during phlebotomy. This study aimed to establish reference intervals for the accurate analysis of VBG results as there is a lack of published data. Method Dr.
View Article and Find Full Text PDFJ Med Biochem
September 2024
University of Verona, Section of Clinical Biochemistry and School of Medicine, Verona, Italy.
Background: Minimizing air aspiration by carefully filling blood gas syringes is crucial to prevent air contamination from causing undesirable variations in gasses and other molecules. While some previous studies investigated this aspect, these are now outdated and only analyzed a limited number of blood gas parameters. Thus, we investigated the effects air contamination in the syringe using a modern blood gas analyzer.
View Article and Find Full Text PDFStomatologiia (Mosk)
December 2024
Northern State Medical University, Arkhangelsk, Russia.
Objective: To study the associations between salivary minerals concentrations and caries experience among adolescents in the conditions of one of the industrial centers of the Far North.
Material And Methods: In total, 146 adolescents aged 15-17 years took part in a cross-sectional study in Syktyvkar, Komi Republic. Caries experience was assessed as DMFT index and its components.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!