Background: Osteochondral injuries have been treated by a variety of methods, each having its own drawbacks. The purpose of this study was to determine the biomechanical feasibility of using a hydrogel implant replacement for an osteochondral core defect. The hypothesis of this study was that the contact pressure of the native knee can be recreated with the use of a hydrogel implant.
Methods: Six cadaver knees were tested in a knee simulator while contact pressures were measured on the tibial plateau. Pressure data was collected in the intact knee, after coring of the condyle and after insertion of a hydrogel implant. Following 1000 gait cycles of fatigue testing, each knee was taken through axial loading indentation testing where the stiffness of the in situ implant was compared to the contralateral condyle.
Findings: While coring significantly reduced the peak pressure at the coring site from 1.8 MPa in the intact knee to 0.0 MPa after coring, implant insertion significantly increased it to 1.2 MPa. There was no significant difference in the peak pressures or the average pressures at the hole location between the intact knee and following implant insertion. After fatigue testing, no macroscopic loosening or implant damage was observed. Based on indentation testing, the stiffness of the medial condyle, 157 N/mm, was significantly less than the lateral condyle, 696 N/mm.
Interpretation: The insertion of the hydrogel implant was able to achieve restoration of contact pressures in the knee supporting the viability of hydrogel implants in the treatment of osteochondral lesions of the knee.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.clinbiomech.2019.04.016 | DOI Listing |
Nat Commun
January 2025
Department of Biomedical Engineering, State University of New York at Binghamton, Binghamton, NY, 13902, USA.
Creating durable, motion-compliant neural interfaces is crucial for accessing dynamic tissues under in vivo conditions and linking neural activity with behaviors. Utilizing the self-alignment of nano-fillers in a polymeric matrix under repetitive tension, here, we introduce conductive carbon nanotubes with high aspect ratios into semi-crystalline polyvinyl alcohol hydrogels, and create electrically anisotropic percolation pathways through cyclic stretching. The resulting anisotropic hydrogel fibers (diameter of 187 ± 13 µm) exhibit fatigue resistance (up to 20,000 cycles at 20% strain) with a stretchability of 64.
View Article and Find Full Text PDFJ Dent Sci
January 2025
Graduate Institute of Clinical Dentistry, School of Dentistry, National Taiwan University and National University Hospital, Taipei, Taiwan.
Background/purpose: Dental implants can restore both function and aesthetics in edentulous areas. However, the absence of cushioning mechanical behavior in implants may limit their clinical performance and reduce the long-term survival rates. This study aimed to establish an implant cushion mechanism that mimicked the natural periodontal ligament, utilizing the properties of composite hydrogels.
View Article and Find Full Text PDFBioact Mater
May 2025
State Key Laboratory for Manufacturing System Engineering, School of Mechanical Engineering, Xi'an Jiaotong University, China.
Implantable neural electrodes are key components of brain-computer interfaces (BCI), but the mismatch in mechanical and biological properties between electrode materials and brain tissue can lead to foreign body reactions and glial scarring, and subsequently compromise the long-term stability of electrical signal transmission. In this study, we proposed a new concept for the design and bioaugmentation of implantable electrodes (bio-array electrodes) featuring a heterogeneous gradient structure. Different composite polyaniline-gelatin-alginate based conductive hydrogel formulations were developed for electrode surface coating.
View Article and Find Full Text PDFInt Med Case Rep J
January 2025
National Scientific Center of Traumatology and Orthopedics Named After Academician Batpenov N.D., Astana, Kazakhstan.
Background: Cartilage defects in the knee joint are areas of damage and wear to the cartilage that normally covers and protects the ends of bones. These defects occur due to sudden injuries, such as trauma or sports accidents, or due to chronic conditions, such as osteoarthritis. Cartilage acts as a shock absorber (cushion absorber), reducing the impact of mechanical stress on the joints, which helps prevent bone damage during movement.
View Article and Find Full Text PDFAdv Sci (Weinh)
January 2025
Medical 3D Printing Center, Orthopedic Institute, Department of Orthopedic Surgery, The First Affiliated Hospital, School of Basic Medical Sciences, MOE Key Laboratory of Geriatric Diseases and Immunology, Suzhou Medical College, Soochow University, Suzhou, Jiangsu, 215000, China.
Stem cell implantation holds promise for enhancing bone repair, but risks of pathogen transmission and malignant cell transformation should not be ignored. Compared to stem cell implantation, recruitment of endogenous stem cells to injured sites is more critical for in situ bone regeneration. In this study, based on the acidic microenvironment of bone injury, an HG-AA-SDF-1α composite hydrogel with a dual-control intelligent switch function is developed by incorporating stromal cell-derived factor (SDF-1α), arginine carbon dots (Arg-CDs), and calcium ions (Ca) into the oxidized hyaluronic acid/gelatin methacryloyl (HG) hydrogel.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!