Objective: This paper focuses on a novel and portable device prototype with optical detectors to quickly and efficiently detect hemozoin (Hz) in blood, aiming at malaria diagnostics.

Methods: Taking advantage of the particular features of malaria parasite in infected blood, particularly the Hz formation, the main innovation described is a portable device for the optical quantification of parasitic Hz in blood, through optical absorbance spectrophotometry. This device comprises detection chambers for fluidic samples, an optical emission and detection system, and a power supply system to provide autonomy. The working principle is based on colorimetric detection, by absorbance, at six specific wavelengths. A detection algorithm relates the absorbance values at all wavelengths to quantify the Hz concentration, thus working as a biomarker of malaria presence and stage.

Results: Under the tested conditions, e.g., in fluidic samples containing synthetic Hz, hemoglobin, and diluted whole blood, the device detected Hz above 1 μg/mL concentrations with 100% sensitivity and 96.3% specificity.

Conclusion: This paper features an autonomous, portable, 1-min analysis time, and low-cost per test device, without the need for samples, centrifugation, allowing the use of whole blood.

Significance: The presented device is a step ahead for meeting the growing clinical demands for reliable, rapid, portable, and quantitative malaria diagnosis.

Download full-text PDF

Source
http://dx.doi.org/10.1109/TBME.2019.2913454DOI Listing

Publication Analysis

Top Keywords

portable device
12
device optical
8
optical quantification
8
diluted blood
8
fluidic samples
8
device
6
portable
5
optical
5
blood
5
quantification hemozoin
4

Similar Publications

Advancements in Wound Bed Preparation of Chronic Wounds.

Surg Technol Int

January 2025

Dr. Phillip Frost Department of Dermatology & Cutaneous Surgery, University of Miami Miller School of Medicine, Miami, Florida.

Chronic wounds are notoriously challenging to heal as they are often halted in their normal healing process. The concept of TIME (Tissue, Inflammation/Infection, Moisture imbalance, Epithelial edge advancement) has been widely utilized in clinical practice to prepare wound beds and promote healing, particularly in longstanding wounds. Traditional methods of wound bed preparation are often inadequate in healing chronic wounds or they may not be tolerated by patients.

View Article and Find Full Text PDF

Hierarchical Porous Aggregate-Enabled Chromatography-Inspired Single-Sensor E-Nose for Volatile Monitoring.

ACS Sens

January 2025

School of Chemistry and Molecular Engineering, In Situ Devices Research Center, Shanghai Key Laboratory for Urban Ecological Processes and Eco-Restoration, East China Normal University, Shanghai 200241, China.

Monitoring volatile organic compounds (VOCs) is crucial for ensuring safety and health. In this study, we introduce a strategy to engineer a chromatography-inspired single-sensor (CISS) e-nose tailored for VOC monitoring. This approach overcomes the limitations of traditional methodologies and conventional e-noses.

View Article and Find Full Text PDF

This study presents a novel optoporation technique using a titanium-coated TiO microstructure (TMS) device activated by an infrared diode laser for highly efficient intracellular delivery. The TMS device, fabricated with 120 nm titanium coating on a titanium dioxide (TiO) microstructure containing microneedles (height ∼2 μm and width ∼4.5 μm), demonstrates enhanced biocompatibility and thermal conductivity compared to the conventional TiO microstructure (MS).

View Article and Find Full Text PDF

Kirkendall Effect-Mediated Transformation of ZIF-67 to NiCo-LDH Nanocages as Oxidase Mimics for Multicolor Point-of-Care Testing of β-Galactosidase Activity and .

Anal Chem

January 2025

Yunnan Key Laboratory of Modern Separation Analysis and Substance Transformation, College of Chemistry and Chemical Engineering, Yunnan Normal University, Kunming 650500, Yunnan Province, P. R. China.

Early and portable detection of pathogenic bacteria is crucial for ensuring food safety, monitoring product quality, and tracing the sources of bacterial infections. Moving beyond traditional plate-culture counting methods, the analysis of active bacterial components offers a rapid means of quantifying bacteria. Here, metal-organic framework (MOF)-derived NiCo-layered double hydroxide nanosheets (LDHs), synthesized via the Kirkendall effect, were employed as highly effective oxidase mimics to generate reactive oxygen species (ROS).

View Article and Find Full Text PDF

A Droplet Microfluidic Sensor for Point-of-Care Measurement of Plasma/Serum Total Free Thiol Concentrations.

Anal Chem

January 2025

Mechanical Engineering, Faculty of Engineering and Physical Sciences, University of Southampton, Southampton SO17 1BJ, U.K.

Total free thiols are an important marker of the whole-body redox state, which has been shown to be associated with clinical outcome in health and disease. Recent investigations have suggested that increased insight may be gained by monitoring alterations of redox state in response to exercise and hypoxia and to monitor redox trajectories in disease settings. However, conducting such studies is challenging due to the requirement for repeated venous blood sampling and intensive lab work.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!