Human motion tracking is widely used for the assessment of movement dysfunction in orthopaedic patients. Currently, most clinical motion analysis centres use marker-based three-dimensional (3D) systems as they are deemed to be the most accurate method. However, due to space, costs and logistics they are not available in many clinical settings. This study compared joint angles measured in functional tests using the novel low-cost Microsoft Kinect Perfect Phorm system with the established marker-based Nexus VICON system. When measuring right and left knee flexion, the average difference between the VICON and Kinect Perfect Phorm measurement was 13.2%, with a of 19.6. Both overestimation and underestimation of the joint angle was recorded in different participants. Although the average percentage difference during hip abduction tests was lower at -3.9%, the range of error was far greater ( = 75). From this, it can be concluded that the level of accuracy presented in the new low-cost Kinect Perfect Phorm system is not yet suitable for clinical assessments. However, for general tests of performance, and for tracking cases where absolute accuracy is less critical, future versions of this software may have a place.

Download full-text PDF

Source
http://dx.doi.org/10.1080/03091902.2019.1599072DOI Listing

Publication Analysis

Top Keywords

kinect perfect
12
perfect phorm
12
joint angle
8
novel low-cost
8
phorm system
8
system
5
validation joint
4
angle measurements
4
measurements comparison
4
comparison novel
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!