Objectives: To compare interfaces performance during home-based automatic titration (APAP).
Methods: Retrospective study based on APAP titration from Obstructive Sleep Apnea Syndrome (OSA) patients.
Results: 707 patients, 513 men (70.6%), were titrated. Masks were 104 pillows (14.7%), group I (GI); 532 nasal (75.2%), group II (GII); and 71 oronasal masks (10%), group III (GIII). We found differences in effective pressure to the device (P90/P95) (GI: 7.13±1.9 vs. GII: 8.3±2.1 vs. GIII: 9.3±2.6 cmHO, p <0.001) but not in final pressure titrated manually (GI: 7.9±1.4 vs. GII: 8.6±1.6 vs. GIII: 9.2±1.9 cm of HO, p >0.5), where lower residual AHI for pillows was p <0.001 and leaks for nasal were p <0.001. No differences were found in compliance (hours) (GI: 6.3±1.2 vs. GII: 6.2±1.1 vs. GIII: 6.1±1.0, p <0.4).
Conclusion: During auto-adjusting titration by CPAP-naïve patients, nasal masks had lower leak rates and nasal pillows presented a similar performance.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6452566 | PMC |
http://dx.doi.org/10.1155/2019/4592462 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!