Comparison of Two Thrombin Generation Methods, CAT and ST-Genesia, in Liver Transplant Patients.

Thromb Haemost

CHU Bordeaux, Laboratoire d'Hématologie - PTRR, Hôpital Pellegrin, Bordeaux, France.

Published: June 2019

Background:  During liver transplantation (LT), thrombin generation (TG) is altered. The most frequently used assay for TG is the Calibrated Automated Thrombogram (CAT). It is designed for series of plasmas and is semi-automated. Complete automation has led to a new device, the ST-Genesia, enabling quantitative standardized TG evaluation.

Objective:  The aim of this observational study was to compare the TG results of the CAT and the ST-Genesia on frozen-thawed plasma samples prepared from the blood of LT patients.

Patients And Methods:  Poor platelet plasma aliquots were prepared from blood samples from six LT patients selected to get the whole range of TG and were assessed with CAT (recombinant human tissue factor [TF] concentration 5 pm) and with ST-Genesia Bleedscreen assay (BS, using 'low' recombinant human TF concentration) and Thromboscreen assay (TS, using 'medium' recombinant human TF concentration). The TG parameters studied were: lag time, peak, time to peak, endogenous thrombin potential, velocity index and start tail.

Results:  BS and TS did not differ significantly from each other whatever the parameter studied, whereas most of the CAT parameters were significantly different from those obtained with BS and TS. Hierarchical clustering analysis of the different parameters of TG showed three homogeneous groups. One cluster gathered TG quantitative parameters from ST-Genesia. A second cluster gathered all the kinetic parameters. The last cluster isolated the quantitative parameters of CAT.

Conclusion:  In patients undergoing LT, TG performed with CAT and with ST-Genesia provided different results, for unknown reasons.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-0039-1685452DOI Listing

Publication Analysis

Top Keywords

cat st-genesia
12
recombinant human
12
thrombin generation
8
prepared blood
8
human concentration
8
time peak
8
cluster gathered
8
quantitative parameters
8
cat
6
st-genesia
6

Similar Publications

Anticoagulant Activity of Nucleic Acid Nanoparticles (NANPs) Assessed by Thrombin Generation Dynamics on a Fully Automated System.

Methods Mol Biol

August 2023

Comprehensive Center for Precision Oncology, Centro de Investigação Translacional em Oncologia (LIM24), Departamento de Radiologia e Oncologia, Faculdade de Medicina da Universidade de São Paulo and Instituto do Câncer do Estado de São Paulo, São Paulo, SP, Brazil.

Rapidly reversible anticoagulant agents have great clinical potential. Oligonucleotide-based anticoagulant agents are uniquely positioned to fill this clinical niche, as they are able to be deactivated through the introduction of the reverse complement oligo. Once the therapeutic and the antidote oligos meet in solution, they are able to undergo isothermal reassociation to form short, inactive, duplexes that are rapidly secreted via filtration by the kidneys.

View Article and Find Full Text PDF

Background: The antiphospholipid syndrome (APS) is classified by the presence of antiphospholipid antibodies (aPL) and thrombotic and/or adverse obstetric outcomes. The diagnosis and risk assessment of APS is challenging. This systematic review investigated if the thrombin generation (TG) assay could be helpful for APS diagnosis and risk assessment.

View Article and Find Full Text PDF

Introduction: The bispecific antibody, emicizumab, is a prophylactic therapy used for the treatment of haemophilia A (HA). Patients may require additional replacement factor VIII (FVIII) to ensure adequate haemostasis. This study investigated the laboratory measurement of severe HA (SHA) plasma spiked with 36 combinations of emicizumab plus recombinant (r) FVIII concentrates.

View Article and Find Full Text PDF

Background: Numerous studies have shown that thrombin generation test (TG) allows a better evaluation of the hemorrhagic and the thrombotic risks. The ST Genesia® is a benchtop, fully automated TG device. However, standardization of the technique and establishment of usual values are essential for its implementation in the routine laboratory.

View Article and Find Full Text PDF

Background: The evaluation of activated protein C (APC) resistance based on the endogenous thrombin potential (ETP) is recommended during the development of steroid contraceptives in women. In 2019, this assay was validated on the calibrated automated thrombogram (CAT) device. However, in view of its screening potential, its automation is essential.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!