Objective: We demonstrate and evaluate the first markerless motion tracker compatible with PET, MRI, and simultaneous PET/MRI systems for motion correction (MC) of brain imaging.

Methods: PET and MRI compatibility is achieved by careful positioning of in-bore vision extenders and by placing all electronic components out-of-bore. The motion tracker is demonstrated in a clinical setup during a pediatric PET/MRI study including 94 pediatric patient scans. PET MC is presented for two of these scans using a customized version of the Multiple Acquisition Frame method. Prospective MC of MRI acquisition of two healthy subjects is demonstrated using a motion-aware MRI sequence. Real-time motion estimates are accompanied with a tracking validity parameter to improve tracking reliability.

Results: For both modalities, MC shows that motion induced artifacts are noticeably reduced and that motion estimates are sufficiently accurate to capture motion ranging from small respiratory motion to large intentional motion. In the PET/MRI study, a time-activity curve analysis shows image improvements for a patient performing head movements corresponding to a tumor motion of ±5-10 mm with a 19% maximal difference in standardized uptake value before and after MC.

Conclusion: The first markerless motion tracker is successfully demonstrated for prospective MC in MRI and MC in PET with good tracking validity.

Significance: As simultaneous PET/MRI systems have become available for clinical use, an increasing demand for accurate motion tracking and MC in PET/MRI scans has emerged. The presented markerless motion tracker facilitate this demand.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6474595PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0215524PLOS

Publication Analysis

Top Keywords

markerless motion
16
motion tracker
16
motion
13
pet mri
12
simultaneous pet/mri
12
motion tracking
8
mri simultaneous
8
pet/mri systems
8
tracker demonstrated
8
pet/mri study
8

Similar Publications

OpenCap, a smartphone-based markerless system, offers a cost-effective alternative to traditional marker-based systems for gait analysis. However, its kinematic measurement accuracy must be evaluated before widespread use in clinical practice. This study aimed to evaluate OpenCap for lower-limb joint angle measurements during walking in patients with knee osteoarthritis (OA) and to compare error metrics between patients and healthy controls.

View Article and Find Full Text PDF

This study aimed to assess the intraday reliability of markerless gait analysis using an RGB-D camera versus a traditional three-dimensional motion analysis (3DMA) system with and without a simulated walking assistant. Gait assessments were conducted on 20 healthy adults walking on a treadmill with a focus on spatiotemporal parameters gathered using the RGB-D camera and 3DMA system. The intraday reliability of the RGB-D camera was evaluated using intraclass correlation coefficients (ICC 1, 1), while its consistency with the 3DMA system was determined using ICC (2, 1).

View Article and Find Full Text PDF

Clinical Whole-Body Gait Characterization Using a Single RGB-D Sensor.

Sensors (Basel)

January 2025

German Center for Vertigo and Balance Disorders (DSGZ), LMU University Hospital, LMU Munich, 81377 Munich, Germany.

Instrumented gait analysis is widely used in clinical settings for the early detection of neurological disorders, monitoring disease progression, and evaluating fall risk. However, the gold-standard marker-based 3D motion analysis is limited by high time and personnel demands. Advances in computer vision now enable markerless whole-body tracking with high accuracy.

View Article and Find Full Text PDF

Marker-Less Video Analysis of Infant Movements for Early Identification of Neurodevelopmental Disorders.

Diagnostics (Basel)

January 2025

Research Coordination and Support Service, Istituto Superiore di Sanità, Viale Regina Elena 299, 00161 Rome, Italy.

The early identification of neurodevelopmental disorders (NDDs) in infants is crucial for effective intervention and improved long-term outcomes. Recent evidence indicates a correlation between deficits in spontaneous movements in newborns and the likelihood of developing NDDs later in life. This study aims to address this aspect by employing a marker-less Artificial Intelligence (AI) approach for the automatic assessment of infants' movements from single-camera video recordings.

View Article and Find Full Text PDF

Gait disturbance is one of the most common symptoms in patients with Parkinson's disease (PD) that is closely associated with poor clinical outcomes. Recently, video-based human pose estimation (HPE) technology has attracted attention as a cheaper and simpler method for performing gait analysis than marker-based 3D motion capture systems. However, it remains unclear whether video-based HPE is a feasible method for measuring temporospatial and kinematic gait parameters in patients with PD and how this function varies with camera position.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!