Purpose: To characterize reproducibility of patient breath-hold positioning and compare tracking system performance for Deep Inspiration Breath Hold (DIBH) gated left breast radiotherapy.
Methods: 29 consecutive left breast DIBH patients (655 fractions) were treated under the guidance of Calypso surface beacons with audio-feedback and 35 consecutive patients (631 fractions) were treated using C-RAD Catalyst HD surface imaging with audiovisual feedback. The Calypso system tracks a centroid determined by two radio-frequency transponders, with a manually enforced institutional tolerance, while the surface image based CatalystHD system utilizes real-time biometric feedback to track a pre-selected point with an institutional tolerance enforced by the Elekta Response gating interface. DIBH motion data from Calypso was extracted to obtain the displacement of breath hold marker in ant/post direction from a set-zero reference point. Ant/post point displacement data from CatalystHD was interpreted by computing the difference between raw tracking points and the center of individual gating windows. Mean overall errors were compared using Welsh's unequal variance t-test. Wilcoxon rank sum test were used for statistical analysis with P < 0.05 considered significant.
Results: Mean overall error for Calypso and CatalystHD were 0.33 ± 1.17 mm and 0.22 ± 0.43 mm, respectively, with t-test comparison P-value < 0.034. Absolute errors for Calypso and CatalystHD were 0.95 ± 0.75 mm and 0.38 ± 0.30 mm, respectively, with Wilcoxon rank sum test P-value <2×10. Average standard deviation per fraction was found to be 0.74 ± 0.44 mm for Calypso patients versus 0.54 ± 0.22 mm for CatalystHD.
Conclusion: Reduced error distribution widths in overall positioning, deviation of position, and per fraction deviation suggest that the use of functionalities available in CatalystHD such as audiovisual biofeedback and patient surface matching improves accuracy and stability during DIBH gated left breast radiotherapy.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ejmp.2019.03.025 | DOI Listing |
Cancers (Basel)
January 2025
Department of Radiation Oncology, Miami Cancer Institute, Miami, FL 33176, USA.
: Over the past decade, significant advances have been made in image-guided radiotherapy (RT) particularly with the introduction of magnetic resonance (MR)-guided radiotherapy (MRgRT). However, the optimal clinical applications of MRgRT are still evolving. The intent of this analysis was to describe our institutional MRgRT utilization patterns and evolution therein, specifically as an early adopter within a center endowed with multiple other technology platforms.
View Article and Find Full Text PDFInt J Environ Res Public Health
January 2025
Australasian Diving Safety Foundation, Canterbury, VIC 3126, Australia.
As snorkelling and breath-hold diving are conducted in a potentially hostile environment by participants with varying skills and health, fatalities occur. In this study, snorkelling and breath-hold diving fatalities were investigated in Australia from 2000 to 2021 to identify causes and countermeasures. The Australasian Diving Safety Foundation database and the National Coronial Information System were searched to identify snorkelling/breath-hold diving deaths from 2000 to 2021.
View Article and Find Full Text PDFSports (Basel)
January 2025
Motion in Brains Research Group, Centro Superior de Estudios Universitarios La Salle, Universidad Autónoma de Madrid, 28023 Madrid, Spain.
To reduce the risk of syncope, trained breath-hold divers (BHDs) use a specialized breathing technique after surfacing called "hook breathing" (HB). It consists of a full inspiration followed by a Valsalva-like maneuver and with subsequent exhalation performed against resistance to generate continuous positive airway pressure during exhalation. This study analyzed the influence of HB on oxygen saturation recovery after a -40 m depth apnea dive in trained BHDs.
View Article and Find Full Text PDFMetabolites
January 2025
Radiological Sciences, David Geffen School of Medicine at UCLA, Los Angeles, CA 90095, USA.
This study evaluated metabolites and lipid composition in the calf muscles of Type 2 diabetes mellitus (T2DM) patients and age-matched healthy controls using multi-dimensional MR spectroscopic imaging. We also explored the association between muscle metabolites, lipids, and intra-abdominal fat in T2DM. Participants included 12 T2DM patients (60.
View Article and Find Full Text PDFCurr Oncol
January 2025
Department of Radiology, Charité-Universitätsmedizin Berlin, Freie Universität Berlin, Humboldt-Universität zu Berlin and Berlin Institute of Health, Charitéplatz 1, 10117 Berlin, Germany.
Breath-hold T2-weighted half-Fourier acquisition single-shot turbo spin echo (HASTE) magnetic resonance imaging (MRI) of the upper abdomen with a slice thickness below 5 mm suffers from high image noise and blurring. The purpose of this prospective study was to improve image quality and accelerate imaging acquisition by using single-breath-hold T2-weighted HASTE with deep learning (DL) reconstruction (DL-HASTE) with a 3 mm slice thickness. MRI of the upper abdomen with DL-HASTE was performed in 35 participants (5 healthy volunteers and 30 patients) at 3 Tesla.