Evaluation and environmental correction of ambient CO measurements from a low-cost NDIR sensor.

Atmos Meas Tech

Department of Atmospheric and Oceanic Science, University of Maryland, College Park, MD 20742, USA.

Published: January 2017

AI Article Synopsis

  • Non-dispersive infrared (NDIR) sensors are a budget-friendly tool for measuring carbon dioxide levels in the air, but their accuracy isn't always adequate for scientific purposes.
  • A study evaluated six SenseAir K30 NDIR sensors in a lab setting, initially showing RMSE errors ranging from ~5 to 21 ppm compared to a high-precision greenhouse gas analyzer.
  • After calibration and environmental adjustments, the sensors' accuracy improved significantly, reducing the RMSE to between 1.7 and 4.3 ppm, indicating their potential for reliable ambient air monitoring.

Article Abstract

Non-dispersive infrared (NDIR) sensors are a low-cost way to observe carbon dioxide concentrations in air, but their specified accuracy and precision are not sufficient for some scientific applications. An initial evaluation of six SenseAir K30 carbon dioxide NDIR sensors in a lab setting showed that without any calibration or correction, the sensors have an individual root mean square error (RMSE) between ~5 and 21 parts per million (ppm) compared to a research-grade greenhouse gas analyzer using cavity enhanced laser absorption spectroscopy. Through further evaluation, after correcting for environmental variables with coefficients determined through a multivariate linear regression analysis, the calculated difference between the each of six individual K30 NDIR sensors and the higher-precision instrument had an RMSE of between 1.7 and 4.3 ppm for 1 min data. The median RMSE improved from 9.6 for off-the-shelf sensors to 1.9 ppm after correction and calibration, demonstrating the potential to provide useful information for ambient air monitoring.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6463532PMC
http://dx.doi.org/10.5194/amt-10-2383-2017DOI Listing

Publication Analysis

Top Keywords

ndir sensors
12
carbon dioxide
8
sensors
5
evaluation environmental
4
environmental correction
4
correction ambient
4
ambient measurements
4
measurements low-cost
4
ndir
4
low-cost ndir
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!