Background: The aim of this study was to validate the clinical performance of the Beckman Access high-sensitivity cardiac troponin I (hs-cTnI) assay.
Methods: We enrolled patients presenting to the emergency department with symptoms suggestive of acute myocardial infarction (AMI). Final diagnoses were centrally adjudicated by 2 independent cardiologists with all clinical information including cardiac imaging twice: first, using serial hs-cTnT (Elecsys, primary analysis), and second, using hs-cTnI (Architect, secondary analysis) measurements in addition to the clinically used hs-cTn. hs-cTnI Access was measured at presentation and at 1 h. The primary objective was a direct comparison of diagnostic accuracy as quantified by the area under the ROC curve (AUC) of hs-cTnI Access vs the hs-cTnT Elecsys and hs-cTnI Architect assays. Secondary objectives included the derivation and validation of an hs-cTnI Access-specific 0/1-h algorithm.
Results: AMI was the adjudicated final diagnosis in 243 of 1579 (15.4%) patients. The AUC at presentation for hs-cTnI Access was 0.95 (95% CI, 0.94-0.96), higher than hs-cTnI Architect [0.92 (95% CI, 0.91-0.94; < 0.001)] and comparable to hs-cTnT Elecsys [0.94 (95% CI, 0.93-0.95; = 0.12)]. Applying the derived hs-cTnI Access 0/1-h algorithm (derivation cohort n = 686) to the validation cohort (n = 680), 60% of patients were ruled out [sensitivity, 98.9% (95% CI, 94.3-99.8)], and 15% of patients were ruled in [specificity, 95.9% (95% CI, 94.0-97.2)]. Patients ruled out by the 0/1-h algorithm had a survival rate of 100% at 30 days. Findings were confirmed in the secondary analyses by the adjudication including serial measurements of Architect hs-cTnI.
Conclusions: Diagnostic accuracy and clinical utility of the Beckman hs-cTnI Access assay are very high and at least comparable to Roche hs-cTnT and Abbott hs-cTnI assays. NCT00470587.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1373/clinchem.2018.300061 | DOI Listing |
Lancet Digit Health
October 2024
Department of Cardiology, University Heart and Vascular Center, University Medical Center Hamburg-Eppendorf, Hamburg, Germany; University Center of Cardiovascular Science, University Heart and Vascular Center, University Medical Center Hamburg-Eppendorf, Hamburg, Germany; Department for Population Health Innovation, University Heart and Vascular Center, University Medical Center Hamburg-Eppendorf, Hamburg, Germany; German Center for Cardiovascular Research (DZHK), Partner Sites Hamburg/Kiel/Luebeck, Hamburg, Germany. Electronic address:
Background: Point-of-care (POC) high-sensitivity cardiac troponin (hs-cTn) assays have been shown to provide similar analytical precision despite substantially shorter turnaround times compared with laboratory-based hs-cTn assays. We applied the previously developed machine learning based personalised Artificial Intelligence in Suspected Myocardial Infarction Study (ARTEMIS) algorithm, which can predict the individual probability of myocardial infarction, with a single POC hs-cTn measurement, and compared its diagnostic performance with standard-of-care pathways for rapid rule-out of myocardial infarction.
Methods: We retrospectively analysed pooled data from consecutive patients of two prospective observational cohorts in geographically distinct regions (the Safe Emergency Department Discharge Rate cohort from the USA and the Suspected Acute Myocardial Infarction in Emergency cohort from Australia) who presented to the emergency department with suspected myocardial infarction.
Diagnostics (Basel)
July 2024
U.O. Clinical Pathology, Medical and Diagnostics Department P.O. Fidenza, Azienda AUSL of Parma, 43125 Parma, Italy.
Background: high-sensitive cardiac TroponinI (hs-cTnI) is widely used for diagnosis of acute coronary syndromes. The latest recommendation for hs-cTnI determination is the protocol 0-1 h finalized to improve the rule out accuracy of the test. A Point of Care Testing able to guarantee these performances could be very useful due to reducing the turnaround time and ruling out patients suspected of ACS, especially by using biological matrices that are not required for centrifuge.
View Article and Find Full Text PDFAnn Clin Lab Sci
May 2024
Department of Pathology, The University of Alabama at Birmingham, Birmingham AL, USA
Objective: Cardiac troponin I (cTnI) is one of the most sensitive and specific biomarkers of myocardial injury. The rise and/or fall of cardiac troponins above the 99th percentile of the upper reference limit (URL) is required in the assessment of acute myocardial infarction.
Methods: We analyzed the variation between the Beckman Coulter contemporary cTnI AccuTnI+3 assay and high sensitivity cTnI (hs-cTnI) assay tested on DXI 800 using 424 patient specimens with troponin levels ranging from 0-16754 ng/L.
Am J Physiol Heart Circ Physiol
June 2024
Center for Developmental Health, Knight Cardiovascular Institute, Oregon Health and Science University, Portland, Oregon, United States.
Elevated cardiac troponin I (cTnI), a myocardial damage biomarker, has been reported in cord blood of neonates delivered vaginally or by cesarean section. Although the neonatal peak likely reflects the physiological adjustment to extrauterine life, a better understanding of serial prepartum changes is required to determine physiological causes of fetal cTnI release. We longitudinally sampled eight healthy lambs (20 days before spontaneous birth to 5 days postnatal), and from three fetuses receiving intravenous IGF-1.
View Article and Find Full Text PDFClin Chem Lab Med
June 2024
Section of Clinical Biochemistry, University Hospital of Verona, Verona, Italy.
Objectives: The current study was designed to evaluate the analytical performance of the new Mindray highly sensitive cardiac troponin I (hs-cTnI) chemiluminescent immunoassay on Mindray CL-1200i, as a thorough validation of novel hs-cTnI methods is required before introduction into clinical practice.
Methods: The evaluation of the analytical performance of this hs-cTnI immunoassay encompassed the calculation of the limit of blank (LOB), limit of detection (LOD), functional sensitivity, imprecision, linearity, 99th percentile upper reference limit (URL) and concordance with another previously validated hs-cTnI chemiluminescent immunoassay.
Results: The LOB and LOD were 0.