Objective: People with diabetic peripheral neuropathy (DPN) have impaired gait and balance performance. The aim of this study is to investigate therapeutic effectiveness of mechanical stimulation through a wearable foot compression device equipped in a shoe insole on vibration perception, balance control and gait performance in people with DPN.

Methods: Using a single-arm 4-week intervention study design, we examined effectiveness of daily use of shoes equipped with the foot compression device (Footbeat™, AVEX, Grand Junction, CO, USA) on improving vibration perception threshold (VPT), skin perfusion pressure (SPP), ankle brachial index (ABI), lower extremities edema (circumferences in the calf and ankle), and motor performance (postural sway with eyes open and closed conditions, and gait performance during normal, dual-task and fast walking). Thirty people with type 2 diabetes and symptoms of PN completed the experimental protocol.

Results: Improvements in VPT (before = 27.4 V, after = 23.3 V, = .007, = 0.33, where denotes effect size), center-of-mass sway in the mediolateral direction with both eyes open and closed conditions (before = 0.94 cm, after = 0.76 cm, = .020, = 0.47; before = 1.10 cm, after = 0.83 cm, = .033, = 0.66, respectively), and stride velocity for normal walking, dual-task walking and fast walking tasks (before = 0.87 m/s, after = 0.96 m/s, = .017, = 0.41; before = 0.75 m/s, after = 0.91 m/s, = .001, = 0.77; before = 1.10 m/s, after = 1.20 m/s, = .043, = 0.33, respectively) were found post treatment. There was no significant improvement in SPP, ABI, and circumferences in the calf and ankle.

Conclusions: Our findings suggest the wearable foot compression device may be effective for reducing neuropathic symptoms and enhancing motor performances in people with DPN.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6955449PMC
http://dx.doi.org/10.1177/1932296819839703DOI Listing

Publication Analysis

Top Keywords

foot compression
16
compression device
16
vibration perception
12
mechanical stimulation
8
gait balance
8
people diabetic
8
diabetic peripheral
8
peripheral neuropathy
8
wearable foot
8
gait performance
8

Similar Publications

Fluid Distribution: Response to Intermittent Pneumatic Compression in People With and Without Primary Lymphedema.

Lymphat Res Biol

January 2025

Caring Futures Institute, College of Nursing and Health Sciences, Flinders University, Bedford Park, Australia.

Current understanding of changes in fluid distribution in response to the application of compression in primary lymphedema (PLE) is limited. This study measured fluid distribution before and after one application of standardized intermittent pneumatic compression (IPC) in the lower limbs of people with PLE, compared with those without lymphedema. High-frequency ultrasound (HFU) was used to measure dermal fluid, bioimpedance to measure segmental fluid, and percent water content (PWC) to measure fluid at specific anatomical points.

View Article and Find Full Text PDF

A Novel Dynamic Growth Rod Inducing Spinal Growth Modulation for the Correction of Spinal Deformities.

JOR Spine

March 2025

Beijing Key Laboratory for Design and Evaluation Technology of Advanced Implantable & Interventional Medical Devices, Beijing Advanced Innovation Center for Biomedical Engineering, School of Biological Science and Medical Engineering Beihang University Beijing China.

Background: Growth rods are the gold standard for treating early-onset scoliosis (EOS). However, current treatments with growth rods do not optimize spinal growth in EOS patients, and frequent distraction surgeries significantly increase complications, imposing considerable economic and psychological burdens on patients. An improved growth rod is urgently required to address the need for dynamic growth and external regulation.

View Article and Find Full Text PDF

O-Arm Navigation-Guided Unilateral Biportal Endoscopic Decompression of Far-Out Syndrome.

Neurospine

December 2024

Department of Neurosurgery, Spine Center, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Korea.

The main aim of this video article is to demonstrate the combined use of O-arm navigation and unilateral biportal endoscopy (UBE) to manage far-out syndrome (FOS). In FOS there is entrapment and compression of the fifth lumbar nerve beyond the foramen and between L5 transverse process and the sacral ala at the lumbosacral junction. Conventional microscopic decompression using a paraspinal approach had been the gold standard for its management.

View Article and Find Full Text PDF

Biodegradable PHBVHHx-PEG/Collagen Hydrogel Scaffolds for Cartilage Repair.

Tissue Eng Part A

January 2025

Department of Orthopedic Surgery and Orthopedic Research Institute, Sports Medicine Center, West China Hospital, Sichuan University, Chengdu, China.

Recently, there has been increased attention on the treatment of cartilage repair. Overall, we constructed PHBVHHx-COL, a composite hydrogel of PHBVHHx-co-PEG and collagen, and evaluated its cartilage repair efficacy through and studies using hydrogel loaded with peripheral blood-derived mesenchymal stem cells (PBMSCs). Rheological properties and compressive mechanical properties of the hydrogels were systematically evaluated.

View Article and Find Full Text PDF

Popliteal artery entrapment syndrome has congenital and functional causes. It mostly affects young people. There are six types of popliteal artery entrapment syndrome.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!