Measurement precision and accuracy of high artificial enrichment N and C tracer samples.

Rapid Commun Mass Spectrom

Institute of Agricultural Sciences, ETH Zurich, Universitätstrasse 2, 8092, Zurich, Switzerland.

Published: July 2019

Rationale: Oversaturation of the Faraday cup amplifiers of isotope ratio mass spectrometers when using tracers that are highly enriched in heavier isotopes (up to 99.9%) remains a major bottleneck to obtaining high-precision measurements. The memory effect plays a key role in reducing tracer sample measurement precision and accuracy. Several sample preparation approaches are known to reduce memory effects and to improve tracer sample measurement precision. However, the potential benefits when using very high enrichment tracer samples (> +1000 mUr) have not been tested.

Methods: In this study, we test how specific sample positioning for measurements and frequent use of natural isotope abundance reference materials within the sequence affects the precision and accuracy of isotopic ratio analyses when using a Flash elemental analyser coupled to a Delta XP isotope ratio mass spectrometer for very high enrichment (> +22000 mUr) N tracer sample measurements. Furthermore, we investigate if tracer sample dilution with natural isotope abundance materials reduces memory effects and increases measurement precision and accuracy when measurements of high-enrichment N and C biomass tracer samples are conducted.

Results: Frequent use of natural isotope abundance materials and specific positioning increased N tracer sample precision, but it had a negative effect on the precision of quality control substances. N and C tracer sample dilution improved measurement precision by a maximum of ±0.9 mUr; however, a strong linear relationship between the original and the calculated φ values was found. Highly enriched N tracer samples caused a maximum memory effect of 0.11%. High levels of N abundance within the samples affected measurement accuracy by an average of 6.7%.

Conclusions: We conclude that highly enriched tracer samples do not require dilution before analysis. Tracer sample precision can be improved by using a specific measurement order of expected isotope abundance and by the frequent use of natural abundance reference materials.

Download full-text PDF

Source
http://dx.doi.org/10.1002/rcm.8451DOI Listing

Publication Analysis

Top Keywords

tracer sample
28
measurement precision
20
tracer samples
20
precision accuracy
16
isotope abundance
16
tracer
12
highly enriched
12
frequent natural
12
natural isotope
12
sample
9

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!