Background: Transepidermal water loss (TEWL) and surface capacitance measure skin barrier permeability and stratum corneum (SC) hydration, respectively, and are frequently utilized in atopic dermatitis clinical trials. Many barrier devices are costly and often used only in the academic setting. GPSkin is a low-cost, patient-operated device that measures both TEWL and SC hydration. This study aimed to test the reliability of GPSkin and assess its correlation with current industry standards.

Materials And Methods: GPSkin was compared to the Biox AquaFlux (TEWL) and Courage-Khazaka Corneometer (SC hydration). Participants with healthy skin (n = 50) collected measurements with GPSkin in Trial 1 without any device education and in Trial 2 with additional instruction. In Trial 2, the investigator also performed measurements with GPSkin. Spearman's coefficients (r ) were performed to assess device correlation. Intraclass correlation coefficients (ICC) were calculated to determine reliability.

Results: Overall, GPSkin was moderately correlated with current industry device measurements for TEWL (Trial 1 r :0.48; Trial 2 r :0.40 participant, 0.34 investigator) and SC hydration (Trial 1 r :0.63; Trial 2 r :0.45). GPSkin demonstrated "good" test-retest reliability for both TEWL (ICC: 0.89) and SC hydration (ICC: 0.85) measurements when participants were provided with some device education. There was no difference in reliability between participants provided with device education and investigators.

Conclusion: Based on these findings, we concluded that GPSkin provides reasonably precise and reliable measurements of SC hydration and TEWL as compared to current devices.

Download full-text PDF

Source
http://dx.doi.org/10.1111/srt.12692DOI Listing

Publication Analysis

Top Keywords

device education
12
skin barrier
8
gpskin
8
current industry
8
measurements gpskin
8
participants provided
8
provided device
8
device
7
trial
7
tewl
6

Similar Publications

To evaluate the accuracy of home self-monitoring portable blood glucose meters, we analyzed the current problems of patients using portable blood glucose meters and put forward reasonable suggestions. A self-designed questionnaire was used to survey 142 patients and 132 healthcare professionals. The questionnaire consisted of 16 items with an overall score ranging from 1 to 13 (with a higher score indicating better experience).

View Article and Find Full Text PDF

Background: This study investigates the therapeutic efficacy of dynamic neuromuscular stabilization (DNS) technology paired with Kinesio Taping in patients with persistent nonspecific low back pain, as well as the effect on neuromuscular function and pain self-efficacy.

Methods: A randomized controlled clinical study was conducted to collect clinical data on DNS combined with KT for the treatment of chronic nonspecific low back pain from November 2023 to April 2024. The inclusion criteria were patients with chronic nonspecific lower back pain, aged between 18 and 30 years old, and without serious underlying medical conditions, such as cardiac disease, hypertension, and diabetes.

View Article and Find Full Text PDF

Ultrasensitive Detection of Circulating Plasma Cells Using Surface-Enhanced Raman Spectroscopy and Machine Learning for Multiple Myeloma Monitoring.

Anal Chem

January 2025

Key Laboratory of OptoElectronic Science and Technology for Medicine of Ministry of Education, Fujian Provincial Key Laboratory of Photonics Technology, Fujian Normal University, Fuzhou, Fujian 350117, China.

Multiple myeloma is a hematologic malignancy characterized by the proliferation of abnormal plasma cells in the bone marrow. Despite therapeutic advancements, there remains a critical need for reliable, noninvasive methods to monitor multiple myeloma. Circulating plasma cells (CPCs) in peripheral blood are robust and independent prognostic markers, but their detection is challenging due to their low abundance.

View Article and Find Full Text PDF

Salt-welding strategy for the design of repairable impact-resistant and wear-resistant hydrogels.

Sci Adv

January 2025

School of Materials Science & Chemical Engineering, Ministry of Education Key Laboratory of Impact and Safety Engineering, Ningbo University, Ningbo 315211, China.

Self-healing hydrogels can autonomously repair damage, enhancing their performance stability and broadening their applications as soft devices. Although the incorporation of dynamic interactions enhances self-healing capabilities, it simultaneously weakens the hydrogels' strength. External stimuli such as heating, while accelerating the healing process, may also lead to dehydration.

View Article and Find Full Text PDF

Lubricating Copolymer Brushes Achieving Excellent Antiadhesion and Antibacterial Performance through Hydration and Electrostatic Repulsion Effects.

ACS Appl Mater Interfaces

January 2025

Key Laboratory of High Efficiency and Clean Mechanical Manufacture of Ministry of Education, School of Mechanical Engineering, Shandong University, Jinan 250061, P. R. China.

Interventional catheters have been widely applied in diagnostics, therapeutics, and other biomedical areas. The complications caused by catheter-related bacterial infection, venous thrombosis, and vascular abrasion have become the main reasons for the failure of interventional therapy. In this study, polyacrylamide/poly(acrylic acid) lubricating copolymer brushes were constructed on the surface of catheters and efficiently resisted the adhesion of blood components and bacteria through hydration and electrostatic repulsion effects.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!