Background: Clinical MRI protocols are time-consuming; hence, new faster techniques are needed. One new fast multicontrast MRI technique, called echo planar image mix (EPIMix) (including contrasts T -FLAIR, T -weighted, diffusion-weighted images [DWI], apparent diffusion coefficient [ADC], T *-weighted, and T -FLAIR images) needs to be tested.

Purpose: To assess if EPIMix has comparable diagnostic performance as routine clinical brain MRI.

Study Type: Prospective.

Population: A consecutive series of 103 patients' brain MRI (January 2018 to May 2018).

Field Strength/sequence: 1.5 T or 3T. EPIMix and routine clinical protocol (clinical MRI included all or some of the contrasts T -FLAIR, T -weighted, DWI, T *-weighted, T -FLAIR, 3D-FSE).

Assessment: Two neuroradiologists assessed EPIMix and clinical scans and categorized the images as abnormal or normal and described diagnosis, artifacts, diagnostic confidence image quality, and comparison of imaging time.

Statistical Tests: Pivot tables with diagnostic performance calculated by receiver operating characteristics (ROC) and the area under curve (AUC). Disease categorization and image quality measures were evaluated. The study protocol is published at ClinicalTrials.gov NCT03338270.

Results: After exclusion of 21 patients, 82 patients had a routine clinical MRI with comparable contrasts to EPIMix and were evaluated. The diagnostic performance to categorize a full brain MRI investigation as abnormal or normal was comparable between EPIMix (AUC 0.99 (95% confidence interval [CI] 0.97-1.00) and 0.99 (95% CI 0.97-1.00)) and routine clinical MRI (n = 82). Sensitivity was 95% (95% CI 88-95) and 93% (95% CI 86-98), and specificity 100% (95% CI 97-100) and 100% (95% CI 90-100). Disease categorization was congruent between EPIMix and clinical routine MRI in 90% (reader 2) and 93% (reader 1). Image quality was generally rated lower for EPIMix (P < 0.001). Imaging time was 78 seconds for EPIMix and for the same contrasts 12 minutes 29 seconds for conventional 3T MRI.

Data Conclusion: EPIMix has comparable diagnostic performance (disease identification and categorization) for most patients investigated in clinical routine. Level of Evidence 2 Technical Efficacy: Stage 2 J. Magn. Reson. Imaging 2019;50:1824-1833.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jmri.26742DOI Listing

Publication Analysis

Top Keywords

diagnostic performance
20
clinical mri
16
routine clinical
16
image quality
12
epimix
11
clinical
9
full brain
8
mri
8
contrasts -flair
8
-flair -weighted
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!