On using Rayleigh scattering for in situ spectral calibration of Thomson scattering diagnostics.

Rev Sci Instrum

Max-Planck-Institut für Plasmaphysik, D-17491 Greifswald, Germany.

Published: March 2019

A new method for in situ spectral calibration of Thomson scattering diagnostics is proposed. The idea of the method is to apply a wavelength tunable optical parametric oscillator for measurements of Rayleigh scattering at different wavelengths, from which relative sensitivities can be calculated. This extends the usual approach where Rayleigh scattering is used only at a single wavelength for the absolute calibration and spectral sensitivities are obtained separately. With the new approach, the full diagnostic setup is spectrally calibrated at once. Such a calibration can be repeated at regular intervals during an experimental campaign since it does not require a break of the vacuum. In this paper, the Rayleigh scattering calibration is tested in a laboratory setup with a sample Wendelstein 7-X (W7-X) polychromator. It is shown that relative sensitivities of spectral channels can be recovered with a sufficient resolution even under conditions of significant stray light. The stray light is overcome by measuring the linear dependence of the scattered signal on the gas pressure. Good results of laboratory tests motivate the installation of the new calibration system for the Thomson scattering diagnostic at W7-X.

Download full-text PDF

Source
http://dx.doi.org/10.1063/1.5072781DOI Listing

Publication Analysis

Top Keywords

rayleigh scattering
16
thomson scattering
12
situ spectral
8
spectral calibration
8
calibration thomson
8
scattering diagnostics
8
relative sensitivities
8
stray light
8
calibration
6
scattering
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!