Accuracy of five plasma calibrated glucometers to screen for and diagnose gestational diabetes mellitus in a low resource clinic setting.

J Clin Transl Endocrinol

MRC Developmental Pathways to Health Research Unit, Department of Paediatrics, Faculty of Health Sciences, University of the Witwatersrand, York Road, Parktown, Johannesburg, South Africa.

Published: June 2019

Aims: We investigated the clinical and analytic accuracy of five plasma calibrated glucometers, the use of which is advocated by the World Health Organisation and the International Federation of Gynaecology and Obstetrics, to screen for and diagnose gestational diabetes mellitus (GDM) in low resource settings.

Methods: 592 consecutive black African women underwent a 75 g oral glucose tolerance test (OGTT) at 24-28 weeks gestation at an urban South African community health clinic. Capillary glucose was measured by one of five glucometer brands, each paired with a routine laboratory hexokinase method of plasma glucose measurement. The laboratory results served as the gold standard reference test for GDM diagnosis. World Health Organisation GDM diagnostic thresholds were applied to glucometer and laboratory results.

Results: Glucometer and laboratory determined GDM prevalence was 75/592 (12.7%) and 30/592 (5.1%) with an elevated fasting glucose diagnostic in 64/75 (85%) and 24/30 (80%) of cases respectively. The proportion of glucometer results fulfilling ISO 15197:2013 recommended analytic accuracy at fasting, 60, and 120 min of the OGTT was 92.4%, 49.8% and 61.5%, with Bland Altman method revealing a positive glucometer bias of 0.22 mmol/l (-0.69-1.12 mmol/l), 0.96 mmol/l (-0.65-2.56 mmol/l) and 0.73 mmol/l (-0.73-2.19 mmol/l) respectively. Only three of the glucometer brands evaluated fulfilled ISO 15197:2013 analytic accuracy requirements and this was only achieved at fasting. All glucometers tested were inaccurate at one and two hours of the OGTT.

Conclusions: Not all glucometers may be suitable for GDM screening as only three were accurate compared to the reference test and then only at fasting of the OGTT. Importantly, laboratory fasting glucose was diagnostic of GDM in 80% of cases in this study population.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6424057PMC
http://dx.doi.org/10.1016/j.jcte.2018.12.003DOI Listing

Publication Analysis

Top Keywords

analytic accuracy
12
accuracy plasma
8
plasma calibrated
8
calibrated glucometers
8
screen diagnose
8
diagnose gestational
8
gestational diabetes
8
diabetes mellitus
8
low resource
8
health organisation
8

Similar Publications

A neuromechanics solution for adjustable robot compliance and accuracy.

Sci Robot

January 2025

Research Center for Information and Communication Technologies, Department of Computer Engineering, Automation and Robotics, University of Granada, Granada, Spain.

Robots have to adjust their motor behavior to changing environments and variable task requirements to successfully operate in the real world and physically interact with humans. Thus, robotics strives to enable a broad spectrum of adjustable motor behavior, aiming to mimic the human ability to function in unstructured scenarios. In humans, motor behavior arises from the integrative action of the central nervous system and body biomechanics; motion must be understood from a neuromechanics perspective.

View Article and Find Full Text PDF

Comment on "Physician vs. AI-generated messages in urology: evaluation of accuracy, completeness, and preference by patients and physicians".

World J Urol

January 2025

Department of Research Analytics, Saveetha Dental College and Hospitals, Saveetha Institute of Medical and Technical Sciences, Saveetha University, Chennai, India.

View Article and Find Full Text PDF

Innovative Infrared Spectroscopic Technologies for the Prediction of Deoxynivalenol in Wheat.

ACS Food Sci Technol

January 2025

Institute of Analytical and Bioanalytical Chemistry, Ulm University, Albert-Einstein-Allee 11, Ulm 89075, Germany.

Mycotoxin contamination in cereals is a global food safety concern. One of the most common mycotoxins in grains is deoxynivalenol (DON), a secondary metabolite produced by the fungi and . Exposure to DON can lead to adverse health effects in both humans and animals including vomiting, dizziness, and fever.

View Article and Find Full Text PDF

The current research presents novel LC-TQ-MS/MS and cost-effective UPLC methods intended for the accurate quantification of mefenamic acid-N-nitroso drug substance-related impurity-NDSRI (N-MFA) in mefenamic acid (MFA) tablet and pediatric suspension dosage forms. The acceptable intake of N-MFA is derived from the TD50 (Median Toxic Dose-50%) value of N-nitroso diphenylamine. The analytical separation was achieved for the UPLC method using an XBridge BEH Shield RP18 Column (150 × 3.

View Article and Find Full Text PDF

Objective: Post-surgical lip symmetry assessment is a key indicator of cleft repair success. Traditional methods rely on distances between anatomical landmarks, which are impractical for video analysis and overlook texture and appearance. We propose an artificial intelligence (AI) approach to automate this process, analyzing lateral lip morphology for a quantitative symmetry evaluation.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!