Introduction: Clinical trials that involve participants from multiple sites necessitate standardized and reliable quantitative MRI outcomes to detect significant group differences over time. Metabolite concentrations measured by proton MRS ( H-MRS) provide valuable information about in vivo metabolism of the central nervous system, but can vary based on the acquisition and quantitation methods used by different MR sites. Therefore, we investigated the intra- and inter-site reproducibility of metabolite concentrations measured by H-MRS on MRI scanners from a single manufacturer across six sites.

Methods: Five healthy controls were scanned twice within 24 h at six participating 3 T MR sites with large single-voxel PRESS (TE/TR/NSA = 36 ms/4000 ms/56) and anatomical images for voxel positioning and correction of partial volume relaxation. Absolute metabolite concentrations were calculated relative to the T and T relaxation corrected signal from water. Intra- and inter-site reproducibility was assessed using Bland-Altman plots and intra- and inter-site coefficient of variation (CoV) as well as intra- and inter-site intra-class correlation coefficient.

Results: The median intra-site CoVs for the five major metabolite concentrations ([NAA], [tCr], [Glu], [tCho] and [Ins]) were between 2.5 and 5.3%. Inter-site CoVs were also low, with the median CoVs for all metabolites between 3.7 and 6.4%. Metabolite concentrations were robust to small inconsistencies in voxel placement and site was not the driving factor in the variance of the measurement of any metabolite concentration. Between-subject differences accounted for the majority of the concentration variability for creatine, choline and myo-inositol (42-65% of the variance).

Conclusion: A large single-voxel H-MRS acquisition from a single manufacturer's MRI scanner is highly reproducible and reliable for multi-site clinical trials.

Download full-text PDF

Source
http://dx.doi.org/10.1002/nbm.4083DOI Listing

Publication Analysis

Top Keywords

intra- inter-site
20
metabolite concentrations
20
inter-site reproducibility
12
clinical trials
8
concentrations measured
8
large single-voxel
8
metabolite
6
intra-
5
concentrations
5
inter-site
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!