Purpose: The aim of the present study was to compare the positional accuracy of varian's exact-arm (E-arm) and retractable-arm (R-arm) supporting electronic portal imaging device (EPID) systems (amorphous silicon flat-panel detector) using the intensity-modulated radiotherapy (IMRT) graticule phantom.

Materials And Methods: The known shifts of 0.5, 1.0, and 1.5 cm were introduced to the given phantom in longitudinal, lateral, and vertical directions, respectively, with respect to treatment couch of medical linear accelerator. The experiment was repeated for different gantry angle and varying source to imager distances (SIDs). The images were acquired for each shift at varying SIDs and beam orientations for both EPID supporting systems. The corresponding shifts obtained from treatment planning system (TPS) were recorded and compared.

Results: The known (expected) and observed (recorded from TPS) shifts obtained for different beam angles (namely, 0°, 90°, 180°, and 270° for anterior, left lateral, posterior, and right-lateral portal images, respectively) in the longitudinal, lateral, and vertical direction at varying SID were compared. The maximum shift in the observed value from the expected one was 3 and 2 mm, respectively, out of the all beam configuration for R-arm and E-arm. These shifts were randomly observed for all imager position and beam orientation.

Conclusion: The IMRT graticule phantom is an effective tool to check the mechanical characteristic and consistency of different EPID supporting arms. The effect of EPID sag due to gravity (gantry and treatment couch) was not significant for detection of shift in patient's position. The E-arm support EPID has better mechanical stability and accuracy in detection of patient's position than that of R-arm.

Download full-text PDF

Source
http://dx.doi.org/10.4103/jcrt.JCRT_707_17DOI Listing

Publication Analysis

Top Keywords

positional accuracy
8
accuracy varian's
8
varian's exact-arm
8
electronic portal
8
portal imaging
8
imaging device
8
intensity-modulated radiotherapy
8
graticule phantom
8
imrt graticule
8
longitudinal lateral
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!