X-ray examination dose surveys: how accurate are my results?

Eur Radiol

Department of Radiology, EPICURA, Clinique Louis Caty, Rue Louis Caty 136, 7331, Baudour, Belgium.

Published: October 2019

Objectives: To determine the variabilities of dose-area-products (DAP) of frequent X-ray examinations collected for comparison with diagnostic reference levels (DRLs).

Methods: DAP values of chest, abdomen, and lumbar spine examinations obtained on devices from two manufacturers were collected in three centers over 1 to 2 years. The variability of the average DAP results defined as the 95% confidence interval in percentage of their median value was calculated for increasing sample sizes, each examination and center. We computed the sample sizes yielding variabilities lower or equal to 25% and 10%. The effect of narrowing patient selection based on body weight was also investigated (ranges of 67-73 Kg, or 60-80 Kg).

Results: DAP variabilities ranged from 75 to 170% of the median value when collecting small samples (10 to 20 DAP). To reduce this variability, larger samples are needed, collected over up to 2 years, regardless of the examination and center. A variability ≤ 10% could only be reached for chest X-rays, requiring up to 800 data. For the abdomen and lumbar spine, the lowest achievable variability was 25%, regardless of the body weight selection, requiring up to 400 data.

Conclusion: Variabilities in DAP collected through small samples of ten data as recommended by authorities are very high, but can be reduced down to 25% (abdomen and lumbar spine) or even 10% (chest) through a substantial increase in sample sizes. Our findings could assist radiologists and regulatory authorities in estimating the reliability of the data obtained when performing X-ray dose surveys.

Key Points: • Low but reasonable variabilities cannot be reached with samples sized as recommended by regulatory authorities. Higher numbers of DAP values are required to reduce the variability. • Variabilities of 10% for the chest and 25% for abdomen and lumbar spine examinations are achievable, provided large samples of data are collected over 1 year. • Our results could help radiologists and authorities interpret X-rays dose surveys.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-019-06055-5DOI Listing

Publication Analysis

Top Keywords

abdomen lumbar
16
lumbar spine
16
sample sizes
12
dose surveys
8
dap values
8
spine examinations
8
examination center
8
body weight
8
small samples
8
reduce variability
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!