An off-beam quartz-enhanced photoacoustic spectroscopy (QEPAS) sensor was designed for ethylene detection using a distributed-feedback quantum cascade laser (QCL) operating in the mid-infrared around 11 μm. The acoustic microresonator configuration was experimentally optimized using an original open-cell photoacoustic setup with a MEMS microphone. Correction factors based on theoretical acoustic models were introduced in order to accurately describe the response of millimeter-sized acoustic resonators. The optimized QEPAS sensor exhibited a limit of detection of 60 ppb for 60 s integration, giving a NNEA of 4.8 × 10 W·cm·Hz.