In this study, we developed the prototype of an optical imaging-based point-of-care (POC) device for monitoring human immunodeficiency virus (HIV)/acquired immunodeficiency syndrome (AIDS) progression that can detect CD4+ T-lymphocytes in human blood. The proposed portable cell-counting system, Helios CD4 Analyzer (Helios), can acquire sample images and analyze the cells automatically using a simple fluorescence imaging module and sample cartridge with a three-dimensional (3D) helical minichannel. The helical minichannel formed on the cylindrical surface enables the sample cartridge to hold a cell suspension present in a fixed sample volume for absolute counting of the cells. With a given total channel length, the helical minichannel-based sample cartridge is smaller than the conventional sample cartridge with a planar microchannel. The implemented nut and bolt mechanism allows the scanning of a relatively large volume of the sample along the helical minichannel by just rotating the cylindrical chamber coupled with a single DC motor rather than using a two-axis motorized translation stage, which considerably simplifies the associated electromechanical parts. It has distinct advantages over the existing devices because of its small size and simple scanning mechanism. We optimized various imaging parameters to enhance the fluorescence detection efficiency of the prototype. Performance evaluations using human blood samples demonstrated good agreement for low CD4 count between the Helios and the PIMA, one of the most widely used POC CD4+ analyzers.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6466421PMC
http://dx.doi.org/10.3390/bioengineering6010024DOI Listing

Publication Analysis

Top Keywords

helical minichannel
16
sample cartridge
16
nut bolt
8
human blood
8
sample
7
helical
5
bolt microfluidics
4
microfluidics helical
4
minichannel
4
minichannel counting
4

Similar Publications

The CD4 (cluster of differentiation 4) counting method is used to measure the number of CD4+ T-lymphocytes per microliter of blood and to evaluate the timing of the initiation of antiretroviral therapy as well as the effectiveness of treatment in patients with human immunodeficiency virus. We developed a three-dimensional helical minichannel-based sample cartridge in which a thread-like microgroove formed in the cylindrical surface and configured a particle-positioning and imaging system equipped with a single DC (direct current) motor that can be controlled by a smartphone application. Confinement and enrichment of CD4 cells within a sharp focal depth along the helical minichannel is accomplished by spinning the cylindrical sample cartridge at high speed before acquiring cell images and thus CD4+ cells with weak fluorescence intensity can be detected even in a channel much deeper than existing two-dimensional flat chambers without an autofocusing module.

View Article and Find Full Text PDF

Nut and Bolt Microfluidics with Helical Minichannel for Counting CD4+ T-Cells.

Bioengineering (Basel)

March 2019

Department of Mechanical Engineering, Graduate School, Kookmin University, 77 Jeongneung-ro, Seongbuk-gu, Seoul 02707, Korea.

In this study, we developed the prototype of an optical imaging-based point-of-care (POC) device for monitoring human immunodeficiency virus (HIV)/acquired immunodeficiency syndrome (AIDS) progression that can detect CD4+ T-lymphocytes in human blood. The proposed portable cell-counting system, Helios CD4 Analyzer (Helios), can acquire sample images and analyze the cells automatically using a simple fluorescence imaging module and sample cartridge with a three-dimensional (3D) helical minichannel. The helical minichannel formed on the cylindrical surface enables the sample cartridge to hold a cell suspension present in a fixed sample volume for absolute counting of the cells.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!