Introduction: Although from a technological point of view, progress is impressive, most implantable hearing devices for conductive or mixed hearing loss have a limited capacity. These devices all bypass the impaired middle ear; therefore, the desired amplification (gain) should be based on the cochlear hearing loss (component) only. The aim of the study is to review the literature with regard to accomplished gain with current implantable devices.

Method: Thirty-one articles could be included. Aided thresholds were compared with prescribed values, based on cochlear hearing loss (bone-conduction thresholds), according to the well-validated NAL rule.

Results: For the majority of the studies, NAL targets were not met. Variation in accomplished gain between implant teams was unacceptably large, largely independent of the type of device that was used. NAL targets were best met at 2 kHz, with worse results at the other frequencies.

Conclusion: Large variations in reported results were found, which primarily depended on implant center. Based on the analyses, a pragmatic fitting procedure is proposed which should minimize the differences between implant centres.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6795533PMC
http://dx.doi.org/10.1097/MAO.0000000000002183DOI Listing

Publication Analysis

Top Keywords

hearing loss
16
conductive mixed
8
mixed hearing
8
implant center
8
based cochlear
8
cochlear hearing
8
accomplished gain
8
nal targets
8
hearing
5
efficacy auditory
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!