Background: Extracorporeal membrane oxygenation constitutes a complex support modality, and accurate monitoring is required. An ideal monitoring system should promptly detect ECMO malfunctions and provide real-time information to optimize the patient-machine interactions. We tested a new volumetric capnometer which enables continuous monitoring of membrane lung carbon dioxide removal (V'COML), to help in estimating the oxygenator performance, in terms of CO removal and oxygenator dead space (VDsML).

Methods: This study was conducted on nine pigs undergoing veno-arterial ECMO due to cardiogenic shock after induced acute myocardial infarction. The accuracy and reliability of the prototype of the volumetric capnometer (CORESET, by Eurosets srl, Medolla, Italy) device was evaluated for V'COML and VDsML measurements by comparing the obtained measurements from the new device to a control capnometer with the sweep gas values. Measurements were taken at five different levels of gas flow/blood flow ratio (0.5-1.5). Agreement between the corresponding measurements was taken with the two methods. We expected that 95% of differences were between d - 1.96s and d + 1.96s.

Results: In all, 120 coupled measurements from each device were obtained for the V'COML calculation and 40 for the VDsML. The new capnometer mean percentage bias (95% confidence interval limits of agreement) was 3.86% (12.07-4.35%) for V'COML and 2.62% (8.96-14.20%) for VDsML. A negative proportional bias for V'COML estimation with the new device was observed with a mean of 3.86% (12.07-4.35%). No correlations were found between differences in the coupled V'COML and VDsML measurements and the gas flow/blood flow ratio or temperature. Coupled measurements for V'COML showed strong correlation (r = 0.991; p = 0.0005), as did VDsML calculations (r = 0.973; p = 0.0005).

Conclusion: The volumetric capnometer is reliable for continuous monitoring of CO removal by membrane lung and VDsML calculations. Further studies are necessary to confirm these data.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0267659119833233DOI Listing

Publication Analysis

Top Keywords

volumetric capnometer
16
continuous monitoring
12
membrane lung
12
monitoring membrane
8
lung carbon
8
carbon dioxide
8
dioxide removal
8
v'coml vdsml
8
vdsml measurements
8
measurements device
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!