Rationale: Ion reflections at the quadrupole pre-filter/mass filter boundary lead to the trapping of ions within the pre-filter. The trapped ions lead to space charge effects on the transmission of ions through this region which become more of an issue with the development of high sensitivity mass spectrometers which have bright ions beams.
Methods: Simulations were carried out matching the effective potentials at the pre-filter/mass filter boundary to reduce the field gradients across the boundary by reducing the field radius of the pre-filter. A further rotation of the pre-filter relative to the mass filter was incorporated into the model. Experiments were carried out to reproduce the conditions of the simulations.
Results: Simulations show that matching the effective potentials at the pre-filter/mass filter boundary and rotating the pre-filter by 45° reduced the number of reflected ion trajectories when the mass filter was operated in either the radiofrequency (RF)-only mode or in mass resolving mode. Experiments confirmed the reduced effects of space charge on the transmission of ions through the pre-filter through the measurement of pre-filter transmission profiles with and without the use of an empty step to remove any trapped ions. These results showed significant improvement over the typical pre-filter/mass filter set-up.
Conclusions: Matching the effective potentials at the quadrupole pre-filter/mass filter boundary and rotating the pre-filter result in a pre-filter/mass filter set-up that does not lead to ions that are reflected at the boundary. The modified set-up does not result in the presence of unwanted trapped ions in the quadrupole pre-filter.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/rcm.8428 | DOI Listing |
Rapid Commun Mass Spectrom
May 2019
Sciex, 71 Four Valley Dr., Concord, Ontario, L4K 4V8, Canada.
Rationale: Ion reflections at the quadrupole pre-filter/mass filter boundary lead to the trapping of ions within the pre-filter. The trapped ions lead to space charge effects on the transmission of ions through this region which become more of an issue with the development of high sensitivity mass spectrometers which have bright ions beams.
Methods: Simulations were carried out matching the effective potentials at the pre-filter/mass filter boundary to reduce the field gradients across the boundary by reducing the field radius of the pre-filter.
Rapid Commun Mass Spectrom
May 2019
Sciex, 71 Four Valley Dr., Concord, Ontario, L4K 4V8, Canada.
Rationale: Quadrupole pre-filter transmission profiles are shown to be affected by the presence of trapped ions within the pre-filter of a quadrupole mass spectrometer. Trapped ions are the result of ions reflected at the pre-filter/mass filter boundary and lead to space charge effects that can affect the shape of the pre-filter transmission profile, ion beam stability, peak width and peak intensity. The development of high sensitivity mass spectrometers with bright ion beams leads to greater numbers of trapped ions.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!