Background: In addition to a plethysmograph, Masimo pulse oximeters display a Perfusion Index (PI) value. This study investigated the possible usefulness of PI for monitoring limb tourniquet arterial occlusion.
Methods: Tactical Ratcheting Medical Tourniquets were applied to the thighs of 15 subjects. Tightening ended at one ratchet-tooth advance beyond Doppler- indicated occlusion. The times and pressures of Doppler and PI signal absences and returns were recorded.
Results: Intermittent PI signal error occurred in 149 of 450 runs (PI, 33% versus Doppler, 0%; p < .0001). PI signal loss lagged Doppler-indicated occlusion by 19 ± 15 seconds (mean ± standard deviation, p < .0001). PI Signal Return lagged tourniquet release by 13 ± 7 seconds (Doppler Signal Return took 1 ± 1 seconds following tourniquet release; p < .0001). PI failed to detect early Doppler audible pulse return in 30 of 39 occurrences.
Conclusion: The PI available on Masimo pulse oximeters is not appropriate for monitoring limb tourniquet effectiveness.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.55460/HOAU-RLAW | DOI Listing |