New protocols for coronary computed tomography angiography (CCTA) could lower the radiation dose for patients but influence the image quality. To compare image quality and radiation exposure in step-and-shoot CCTA and high-pitch spiral CCTA. Fifty-nine pairs of patients matched for weight, height, sex and heart rate were included in this study (74 m, 44 f, average age 60 years, age range 29-94 years). Step-and-shoot CCTA and high-pitch spiral CCTA was performed on a third generation dual-source CT in equally sized patient groups. The signal-to-noise ratio (SNR) in the ascending aorta and the coronary arteries were determined for each dataset. Image quality was rated using a five-point scale. We used the t-test for paired samples to compare SNR and effective dose, and the Wilcoxon test to compare image quality scores. Mean effective dose for the step-and-shoot protocol (4.15 ± 3.07 mSv) was significantly higher in comparison to the high-pitch spiral protocol (1.2 ± 0.69 mSv; p < 0.0001). Mean SNR was higher with the step-and-shoot protocol compared to the high-pitch spiral protocol in the aorta, in the left main and peripheral coronary arteries (p < 0.01), in the proximal right coronary artery (p = 0.027). Image quality scores were significantly better for the step-and-shoot protocol (p = 0.0003). Step-and-shoot CCTA has significantly better SNR and overall image quality compared to high-pitch spiral CCTA, but with a mean effective dose more than thrice as high.