Objective:: Laryngeal paralysis of central or peripheral origin can potentially be treated using functional electrical stimulation (FES) of laryngeal muscles. Experiments in canines (dogs) were performed using implant prototypes capable of closed-loop FES to refine engineering designs and specifications, test surgical approaches for implantation, and better understand the in vivo effects of laryngeal muscle stimulation on short- and long-term glottic function.

Study Design:: Prospective, laboratory.

Methods:: We designed and tested a series of microprocessor-based implantable devices that can stimulate glottic opening or closing based on input from physiological control signals (real-time processing of electromyographic [EMG] signals). After acute device testing experiments, 2 dogs were implanted for 8 and 24 months, with periodic testing of closed-loop laryngeal muscle stimulation triggered from EMG signals. In total, 5 dogs were tested for the effects of laryngeal muscle stimulation on vocal fold (VF) posturing in larynges with nerve supplies that were intact (7 VFs), synkinetically reinnervated (2 VFs), or chronically denervated (1 VF). In 3 cases, the stimulation was combined with airflow-driven phonation to study the consequent modulation of phonatory parameters.

Results:: Initial device prototypes used inductive coupling for power and communication, while later iterations used battery power and infrared light communication (detailed descriptions are provided in the Part 1 companion paper). Two animals were successfully implanted with the inductively powered units, which operated until removed at 8 months in 1 animal or for more than 16 months in the second animal. Surgically, the encapsulated implants were well tolerated, and procedures for placing, attaching, and connecting the devices were developed. To simulate EMG control signals in anesthetized animals, we created 2 types of nerve/muscle signal sources. In one approach, a neck muscle had a cuff electrode placed on its motor nerve that was connected to transdermal electrical connection ports for periodic testing. In the second approach, the recurrent laryngeal nerve on one side of the larynx was stimulated to generate a VF EMG signal, which was then used to trigger FES of the paralyzed contralateral side (eg, restoring VF movement symmetry). Implant testing identified effective stimulation parameters and closed-loop stimulation artifact rejection techniques for FES of both healthy and paralyzed VFs. Stimulation levels effective for VF adduction did not cause signs of discomfort during awake testing.

Conclusion:: Our inductive and battery-powered prototypes performed effectively during in vivo testing, and the 2 units that were implanted for long-term evaluation held up well. As a proof of concept, we demonstrated that elicited neck strap muscle or laryngeal EMG potentials could be used as a control signal for closed-loop stimulation of laryngeal adduction and vocal pitch modulation, depending on electrode positioning, and that VFs were stimulable in the presence of synkinetic reinnervation or chronic denervation.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0003489418820545DOI Listing

Publication Analysis

Top Keywords

laryngeal muscle
12
muscle stimulation
12
laryngeal
10
stimulation
9
device testing
8
laryngeal paralysis
8
effects laryngeal
8
control signals
8
periodic testing
8
closed-loop stimulation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!