Introduction: Digital radiography lacks visual clues of exposure techniques used to obtain radiographs, therefore manufacturers have included exposure indicators (EIs). EIs provides feedback about exposure techniques used and evaluating EIs will yield much needed information about exposure trends used in digital radiography.
Methods: A retrospective explorative quantitative study was conducted at nine randomly selected imaging departments in Gauteng, South Africa. Data pertaining to EI was retrospectively collected using quota sampling and compared to manufacturer recommended (MR) standards.
Results: A total of 1422 EIs were collected. 50% of these were within the MR standard. 27% of EI indicated overexposure and 23% indicated underexposure.
Conclusions: Greater evidence of overexposure was noted in the retrospective analysis of the EI. This pilot study shows the need for further investigation into exposure technique practices in digital radiography and the need for measures to halt the evidenced overexposure.