This article describes an improvement of the 3-omega thermal conductivity measurement system facilitating automatic mode measurement. The use of the electrical circuit with negative feedback to compensate the first harmonic in the voltage signal on the heater-thermometer eliminates manual resistance tuning. To test the procedure, measurements of the thermal conductivity of single crystal sapphire AlO were made. The experimental thermal conductivity value is in a good agreement with the reference data.

Download full-text PDF

Source
http://dx.doi.org/10.1063/1.5084103DOI Listing