Aims: To compare transcatheter aortic valve replacement TAVR with self-expandable first-generation Medtronic CoreValve with new-generation Evolut R devices in patients with aortic stenosis.

Methods: Multiple databases were screened for all available reports directly or indirectly comparing CoreValve vs Evolut R. Primary endpoint was device success. Procedural, functional and clinical outcomes were assessed as well.

Results: Ten retrospective series including 12 294 pts. were found. Overall device success rate was 95.5% and was statistically higher in the Evolut R treated patients as compared with CoreValve: 96.6 vs. 94.8%, respectively; RR (risk ratio) 95%CIs (confidence intervals): 1.02 (1.00-1.04); P = 0.01. There were no statistical differences with regard to postoperative mean aortic gradients 8.5 +/- 5.3 vs 7.9 +/- 4.6 with Evolut R and CoreValve. Evolut R valve demonstrated nearly 50% reduction of the risk for moderate-to-severe paravalvilar leak 0.55 (0.39-0.79); P = 0.001; 60% statistically significant lower risk of developing myocardial injury 0.40 (0.22-0.72); P = 0.002 and numerical reductions in the risk of acute kidney injury, vascular complications and bleeding. Together with significantly reduced risk of permanent pacemaker implantation (0.80 [0.67-0.96]; P = 0.02) the above benefits were associated with 40% reduction in the risk of 30-day all-cause mortality with Evolut R as compared to CoreValve: 0.60 (0.37-1.00); P = 0.05.

Conclusions: The use of new-generation Evolut R was associated with improved procedural, functional and clinical outcomes compared with the CoreValve device.

Download full-text PDF

Source
http://dx.doi.org/10.2459/JCM.0000000000000757DOI Listing

Publication Analysis

Top Keywords

compared corevalve
12
aortic valve
8
evolut
8
evolut corevalve
8
new-generation evolut
8
corevalve evolut
8
device success
8
procedural functional
8
functional clinical
8
clinical outcomes
8

Similar Publications

Article Synopsis
  • The study focuses on the impact of tricuspid regurgitation (TR) on heart remodeling, comparing outcomes in patients treated with the TriClip device versus those receiving standard medical therapy in a randomized controlled trial.
  • Researchers utilized advanced imaging techniques, including cardiac magnetic resonance and 4D-CT, to assess heart changes at baseline, 30 days, and one year follow-ups.
  • Results showed that the TriClip significantly reduced TR volume by 70% at 30 days, leading to notable reductions in right ventricular size and area, with these improvements maintained after one year, unlike the control group.
View Article and Find Full Text PDF

Thrombogenic Risk Assessment of Transcatheter Prosthetic Heart Valves Using a Fluid-Structure Interaction Approach.

Comput Methods Programs Biomed

December 2024

Department of Biomedical Engineering, Stony Brook University, Stony Brook, NY 11794, USA. Electronic address:

Background And Objective: Prosthetic heart valve interventions such as TAVR have surged over the past decade, but the associated complication of long-term, life-threatening thrombotic events continues to undermine patient outcomes. Thus, improving thrombogenic risk analysis of TAVR devices is crucial. In vitro studies for thrombogenicity are typically difficult to perform.

View Article and Find Full Text PDF

Native Sinus Hemodynamics and Thrombosis in Transcatheter Heart Valve: Effect of Implant Depth and Coronary Flow.

J Am Heart Assoc

November 2024

Department of Radiology and Research Institute of Radiology, Cardiac Imaging Center University of Ulsan College of Medicine, Asan Medical Center Seoul South Korea.

Background: This study aimed to investigate the hemodynamic and anatomic factors associated with sinus thrombosis following transcatheter aortic valve replacement (TAVR), integrating in vivo patient data analysis and in vitro experiments.

Methods And Results: Postprocedural, 4-dimensional, multiphase computed tomography data from 211 patients enrolled in the ADAPT-TAVR (Anticoagulation Versus Dual Antiplatelet Therapy for Prevention of Leaflet Thrombosis and Cerebral Embolization After Transcatheter Aortic Valve Replacement) study were analyzed. The prevalence of native sinus thrombosis was examined in relation to valve type, implant depth, and anatomic features.

View Article and Find Full Text PDF

Valve-in-valve (ViV) transcatheter aortic valve replacement (TAVR) has been associated with favorable outcomes in patients with degenerated stentless bioprosthesis. However, whether the outcomes after ViV TAVR for failed stentless bioprosthesis differ between balloon-expandable valves (BEVs) and self-expanding valves (SEVs) remains unknown. Therefore, we retrospectively analyzed 59 consecutive patients who underwent ViV TAVR for failed stentless bioprsothesis with BEVs (n = 42) versus SEVs (n = 17) in a single-health care system between 2013 and 2022.

View Article and Find Full Text PDF

Background And Objective: Prosthetic heart valve interventions such as TAVR have surged over the past decade, but the associated complication of long-term, life-threatening thrombotic events continues to undermine patient outcomes. Thus, improving thrombogenic risk analysis of TAVR devices is crucial. In vitro studies for thrombogenicity are typically difficult to perform.