Background: The mechanically expandable Lotus Valve System is a fully repositionable and retrievable valve with an adaptive seal to minimize paravalvular leak (PVL). The aim of this study was to evaluate the short- and long-term safety and efficacy of the new device with focus on a new implantation technique to reduce the need for a permanent pacemaker (PPM) post procedure.
Methods: We performed a prospective single-center, non-randomized evaluation of the Lotus Valve System. The first 100 consecutive Lotus Valve implantations were included in the analysis. Outcome was assessed according to VARC2-criteria. Postoperative pacemaker rates were assessed using the national pacemaker registry and electronic medical records. Mortality at 30 days and 12 months were acquired from the national population registry.
Results: Mean age was 82.7 ± 5.6 years, mean Euroscore I was 25.3 ± 14.5%, mean STS-score was 6.5 ± 4.1% and mean aortic valve area was 0.6 ± 0.1 cm. There were no cases of valve embolization, ectopic valve deployment or additional valve implantation. Device success according to the VARC2-criteria was 97%. The 30-day mortality rate was 3%. Two deaths occurred due to stroke and one due to a ventricular rupture. Major stroke rate was 2% and major vascular complication rate was 2%. The 12-month mortality rate was 14%. At discharge 87% of patients had no/trace PVL, 12% had mild PVL and one patient had a moderate PVL. A total of 13% received a new PPM post valve implantation. Among patients who did not have a PPM before the procedure, the PPM rate was 15.3%.
Conclusions: This single-center evaluation of the Lotus Valve System demonstrated a good clinical outcome with a low mortality, in a high-risk population. Introduction of a new implantation technique resulted in lower PPM rates than previously reported without negatively affecting PVL.
Trial Registration: Current Controlled Trials ISRCTN14952278 , retrospectively registered 06/11/2017.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6390309 | PMC |
http://dx.doi.org/10.1186/s12872-019-1021-7 | DOI Listing |
Gen Thorac Cardiovasc Surg Cases
December 2024
Department of Cardiovascular Surgery, National Cerebral and Cardiovascular Center, Osaka, 564-8565, Japan.
Background: With the rapid expansion of transcatheter aortic valve replacement (TAVR), TAVR valve explantation is also increasing. Nevertheless, previous reports on Lotus Edge valve explantation are limited to only two reports, none of which include intraoperative videos. Therefore, we report the case of an older adult who underwent a 2-year-old Lotus Edge valve explantation, after developing prosthetic valve endocarditis (PVE) and aortic annular abscess, with a strong indication for a TAVR explantation and surgical aortic valve replacement (AVR).
View Article and Find Full Text PDFJACC Cardiovasc Interv
June 2024
Cardiovascular Translational Lab, Centre for Cardiovascular Innovation, Centre for Heart Lung Innovation, University of British Columbia, Vancouver, British Columbia, Canada. Electronic address:
Background: The etiology of transcatheter aortic valve (TAV) degeneration is poorly understood, particularly noncalcific mechanisms.
Objectives: The authors sought to investigate noncalcific and calcific mechanisms of TAV degeneration and evaluate their impact on leaflet function by bench testing, imaging, and histology.
Methods: TAV explants were obtained from the EXPLANT THV registry and clinical institutions.
Glob Cardiol Sci Pract
September 2023
Aswan Heart Centre, Aswan, Egypt.
Introduction: Aortic stenosis (AS) is the leading heart valve disease in developed countries, often caused by calcific degeneration. In low-and-middle-income countries, it's primarily due to RHD. Prevalence of AS increases with age and up to 22.
View Article and Find Full Text PDFJ Clin Med
February 2024
Heart Institute, Medical School, University of Pécs, 7624 Pécs, Hungary.
We report 30-day, 1-year, and 3-year outcomes for a new TAVR programme that used five different transcatheter heart valve (THV) systems. From 2014 to 2020, 122 consecutive patients with severe aortic stenosis (AS) received TAVR based on the Heart Team decision. Outcomes were analysed for the whole study population and in addition the first 63 patients (Cohort A, 2014 to 2019) were compared to the last 59 patients (Cohort B, 2019 to 2020).
View Article and Find Full Text PDFStruct Heart
January 2024
Department of Cardiology and Cardiovascular Research Institute Basel (CRIB), University Hospital Basel, University of Basel, Basel, Switzerland.
The increasing use of transcatheter aortic valves in patients with aortic stenosis has led to a higher number of valve-in-valve procedures due to gradual valve degeneration. We present a case of a 72-year-old woman who received transcatheter aortic valve implantation (TAVI) using a Lotus valve due to severe aortic stenosis, which showed valve degeneration several years after the initial procedure. After heart-team discussion, TAVI-in-TAVI was planned using an Evolut pro+ valve, which allowed for full coverage of the Lotus valve and maintenance of coronary flow, resulting in a well-functioning valve with no regurgitation and normal performance.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!