Self-monitoring of blood glucose is recommended to monitor glycaemic levels for the prevention and treatment of diabetes mellitus.The accuracy of blood glucose monitoring systems (BGMS) directly affects the health of the patients. In this study, we evaluated the accuracy of four different blood glucose monitoring systems, including Gold-Accu Meter, Nova StatStrip Xpress Glucose Hospital Meter, Freestyle Optium Neo H and Accu-Chek Performa according to ISO15197:2003 and ISO15197:2013 criteria. We found that four brands of BGMS (eight BGMS) all met the minimum requirements for system accuracy performance of ISO 15197:2003 and the accuracy criteria of ISO 15197:2013. All results of BGMS lie within zones A and B of the Consensus Error Grid (CEG). In conclusion, all BGMS met the accuracy criteria of ISO 15197:2003 and the accuracy criteria of ISO 15197:2013.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.pcd.2018.12.010 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!