Statement Of Problem: Contamination with salivary fluids or blood during the treatment process changes the preload, remaining torque, and removal torque of retained screws, which ultimately affects the resistance to screw loosening. In previous studies, no consensus has been reached as to whether contamination can be used as a lubricant.
Purpose: The purpose of this 3-dimensional (3D) finite element analysis study was to compare the preload, remaining torque, and removal torque under different frictional conditions in the processes of tightening, waiting period, and removal of abutment screws using a numeric method and finite element analysis.
Material And Methods: Three-dimensional finite element models of a single implant restoration including a crown, dental implant, abutment, and abutment screw, along with the surrounding bone, were constructed. The geometry of all threaded interfaces was designed as a threaded helix. The Abaqus software was used to perform the dynamic simulation of 3 steps such as tightening, waiting period, and removal. Three static and kinetic friction conditions were considered to determine the effect of different frictional conditions. The values of preload at the tightening step, remaining torque at the waiting period, and removal torque at the removal step were evaluated and compared with theoretically predicted values.
Results: The amount of removal torque required to loosen the abutment screw was smaller than the insertion torque for all frictional conditions. By decreasing the coefficient of friction, the remaining torque and the preload increased, and the torsional relaxation and removal torque decreased.
Conclusions: Although the value of the removal torque decreased by decreasing the coefficient of friction, the resistance to screw loosening increased with the increase of the preload and the increase of the remaining torque in the implant complex assembly. Considering the biological complications of fluid lubricants, clinicians may contaminate the abutment screw with lubricants. Gold-coated screws should also be preferred to noncoated screws.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.prosdent.2018.10.035 | DOI Listing |
J Craniofac Surg
December 2024
Department of Oral and Maxillofacial Surgery, College of Dentistry, King Saud University.
Objectives: This study aimed to evaluate the osseointegration properties of titanium bone implants coated with carob-mediated calcium hydroxide nanoparticles biomechanically, radiographically, and histologically on rabbit tibias.
Material And Methods: Forty coated and forty uncoated titanium alloy bone implants were inserted into 20 New Zealand rabbits; each tibia received 2 implants. The rabbits were sacrificed after 4 or 8 weeks, and samples were retrieved for biomechanical evaluation through removal torque test to assess the bond between implant and bone, radiographic evaluation through microcomputed tomography analysis to compare the bone-to-implant contact percentage and bone volume of the peri-implant area, scanning electron microscopic and histologic evaluation through hematoxylin and eosin stain.
Int J Spine Surg
December 2024
Department of Scientific Affairs, Globus Medical, Inc., Audubon, PA, USA.
Background: Anterior column realignment is an attractive minimally invasive treatment for sagittal imbalance. Expandable spacers offer controlled tensioning of the anterior longitudinal ligament (ALL) during release, which could optimize correction and anterior column stability. This study investigated the biomechanical and radiographic effects of single-level anterior-to-psoas lumbar interbody fusion (ATP-LIF) with expandable spacers and sequential ALL release.
View Article and Find Full Text PDFJ Prosthet Dent
November 2024
Associate Professor, State Key Laboratory of Military Stomatology & National Clinical Research Center for Oral Diseases & Shaanxi Engineering Research Center for Dental Materials and Advanced Manufacture, Department of Implant Dentistry, School of Stomatology, The Air Force Medical University, Xi'an, PR China. Electronic address:
Statement Of Problem: Three months after implant placement, difficulty was encountered in unscrewing the healing abutment within the normal torque range (<35 Ncm), resulting in the removal of the implant. Despite all attempts, the implant could not then be separated from the healing abutment. Retrospective analysis of intraoperative photographs revealed the presence of blood or bone debris within the implant before the healing abutment was placed; this was considered to be the likely cause of this phenomenon.
View Article and Find Full Text PDFJ Funct Biomater
November 2024
Department of Basic Sciences, Araçatuba School of Dentistry, São Paulo State University "Júlio de Mesquita Filho", Araçatuba 16018-805, São Paulo, Brazil.
Estrogen deficiency is one of several contributing factors to catabolic changes in bone surrounding dental implants, impairing bone repair in defects requiring bone regeneration. Functionalizing bone substitutes is an alternative approach among various strategies to address this challenge. In this study, the aim was to evaluate the effect of functionalizing deproteinized bovine bone (Bio-Oss, BO) with genistein via sonication on peri-implant bone defects in ovariectomized rats.
View Article and Find Full Text PDFOrthop J Sports Med
November 2024
Department of Orthopaedic Surgery, Stanford University, Stanford, California, USA.
Background: Acetabular labral tear morphology or orientation may influence hip stability.
Hypothesis: A radial tear of the acetabular labrum would result in greater rotational and translational motion compared with a chondrolabral separation.
Study Design: Controlled laboratory study.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!