Background: Pelvic ring-comminuted transforaminal sacral fracture injuries are rotationally and vertically unstable and have a high rate of failure.

Objective: Our study purpose was to use three-dimensional (3D) optical tracking to detect onset location of bone-implant interface failure and measure the distances and angles between screws and line of applied force for correlation to strength of pelvic fracture fixation techniques.

Methods: 3D relative motion across sacral-rami fractures and screws relative to bone was measured with an optical tracking system. Synthetic pelves were used. Comminuted transforaminal sacral-rami fractures were modelled. Each pelvis was stabilised by either (1) two iliosacral screws in S1, (2) one transsacral screw in S1 and one iliosacral screw in S1 and (3) one trans-alar screw in S1 and one iliosacral screw in S1; groups 4-6 consisted of fixation groups with addition of anterior inferior iliac pelvic external fixator. Eighteen-instrumented pelvic models with right ilium fixed simulate single-leg stance. Load was applied to centre of S1 superior endplate. Five cycles of torque was initially applied, sequentially increased until permanent deformation occurred. Five cycles of axial load compression was next applied, sequentially increased until permanent deformation occurred, followed by axial loading to catastrophic failure. A Student  test was used to determine significance ( < 0.05).

Results: The model, protocol and 3D optical system have the ability to locate how sub-catastrophic failures initiate. Our results indicate failure of all screw-based constructs is due to localised bone failure (screw pull-in push-out at the ipsilateral ilium-screw interface, not in sacrum); thus, no difference was observed when not supplemented with external fixation.

Conclusion: Inclusion of external fixation improved resistance only to torsional loading.

Translational Potential Of This Article: Patients with comminuted transforaminal sacral-ipsilateral rami fractures benefit from this fixation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6350021PMC
http://dx.doi.org/10.1016/j.jot.2018.06.005DOI Listing

Publication Analysis

Top Keywords

comminuted transforaminal
8
transforaminal sacral
8
sacral fracture
8
optical tracking
8
sacral-rami fractures
8
screw iliosacral
8
iliosacral screw
8
applied sequentially
8
sequentially increased
8
increased permanent
8

Similar Publications

Comparison of Modified Transiliac Internal Fixators: Mechanical Testing on Pelvic Models.

Acta Chir Orthop Traumatol Cech

November 2021

NTIS - New Technologies for the Information Society, Faculty of Applied Sciences, University of West Bohemia, Plzeň, Czech Republic.

PURPOSE OF THE STUDY Vertically unstable transforaminal sacral fractures can be stabilized with several types of transiliac internal fixators (TIFI): the classical one (TIFI-C), the supraacetabular one (TIFI-A) and by dual application of TIFI (DTIFI). MATERIAL AND METHODS Pelvic models made of solid foam (Sawbones 1301) were used in the study. Mechanical loading tests were performed in order to assess the stiffness of the studied pelvic structures.

View Article and Find Full Text PDF

Objective: To determine whether fully threaded transiliac-transsacral (TI-TS) fixation is biomechanically superior to partially threaded TI-TS fixation of vertically unstable transforaminal sacral fractures.

Methods: Vertically unstable zone 2 sacral fractures were created in 20 human cadaveric pelves with a unilateral osteotomy and resection of 1 cm of bone through the foramen of the sacrum to represent comminution. Ten specimens received either 2 7.

View Article and Find Full Text PDF

Background: Pelvic ring-comminuted transforaminal sacral fracture injuries are rotationally and vertically unstable and have a high rate of failure.

Objective: Our study purpose was to use three-dimensional (3D) optical tracking to detect onset location of bone-implant interface failure and measure the distances and angles between screws and line of applied force for correlation to strength of pelvic fracture fixation techniques.

Methods: 3D relative motion across sacral-rami fractures and screws relative to bone was measured with an optical tracking system.

View Article and Find Full Text PDF

Achieving interfragmentary compression without special drilling technique or screw design.

J Orthop Res

April 2018

Department of Orthopaedic Surgery, Davis Medical Center, University of California, 4860 Y Street, Suite 1700, Sacramento, 95817, California.

Unlabelled: Traditional fracture fixation teaching suggests that fully threaded screws do not provide interfragmentary compression unless placed through a glide hole. Based on this assumption, pelvic surgeons typically use fully threaded screws in the treatment of comminuted transforaminal sacral fractures to limit iatrogenic neuroforaminal stenosis. Clinical experience with fully threaded screws suggests that interfragmentary compression actually does occur.

View Article and Find Full Text PDF

Purpose: Vertically unstable sacral transforaminal fractures can be stabilized with a transiliac internal fixator (TIFI) or two iliosacral screws (IS). This study was designed to compare stiffness between TIFI and IS.

Methods: Using CT images finite element model of the pelvis was developed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!