Background: P-wave terminal force (PTF) > 4000 ms·μV and deep terminal negativity (DTN) are ECG markers of left atrial abnormality associated with both atrial fibrillation and stroke. When the precordial lead V1 is placed higher than the correct position in the fourth intercostal space, it may cause increased PTF and DTN. Several studies have documented that electrode misplacement, especially high placement, is common. The influence of electrode misplacement on these novel ECG markers has not previously been quantified.

Objective: The objective was to assess the influence of electrode misplacement on PTF and DTN.

Method: 12-Lead ECGs were recorded in 29 healthy volunteers from the Department of Cardiology at the Copenhagen University Hospital of Bispebjerg. The precordial electrode V1 was placed in the fourth, third and second intercostal space, giving a total of 3 ECGs per subject. Continuous variables were compared using Dunnett's post-hoc test and categorical variables were compared using Fischer's exact test.

Results: High placement of V1 electrodes resulted in a more than three-fold increase of PTF (IC4 = 2267 ms·μV, IC2 = 7996 ms·μV, p-value < 0.001). There was a similar increase of DTN (IC4 = 0%, IC2 = 28%, p-value < 0.001). P-wave area and amplitude of the negative deflection increased, and P-wave area and amplitude of the positive deflection decreased. The P-wave shape changed from being predominantly positive or biphasic in IC4 to 90% negative in IC2. The PR-duration and P-wave duration were not altered by electrode placement.

Conclusion: High electrode placement results in significant alteration of PTF and DTN in lead V1.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jelectrocard.2019.01.088DOI Listing

Publication Analysis

Top Keywords

electrode misplacement
12
p-wave terminal
8
terminal force
8
ecg markers
8
intercostal space
8
high placement
8
influence electrode
8
variables compared
8
common source
4
source miscalculation
4

Similar Publications

Objective To develop an algorithm, based on the voltage matrix, for detecting regular cochlear implant (CI) electrode position during the implantation procedure, tip fold-over or basal kinking for lateral-wall electrodes. The availability of an algorithm would be valuable in clinical routine, as incorrect positioning of the electrode array can potentially be recognized intraoperatively. Design In this retrospective study intraoperative voltage matrix and postoperative digital volume tomography of 525 CI recipients were analyzed.

View Article and Find Full Text PDF

Deep brain stimulation (DBS) is a neurosurgical procedure that depends on high-accuracy targeting of structures to implant electrodes within the brain. The positioning of these electrodes in the brain determines the long-term efficacy of treating diseases such as Parkinson's disease, essential tremor, or dystonia. Misplaced electrodes in DBS can lead to poor efficacy and stimulation-induced side effects.

View Article and Find Full Text PDF

Introduction: Cardiac implantable electronic device (CIED) complications present significant challenges in clinical practice, especially in elderly patients with multiple comorbidities. Common adverse events include infection, lead malfunction, and device migration. Twiddler's Syndrome, a rare but serious CIED complication characterised by patient manipulation causing lead displacement and device malfunction, is often underreported.

View Article and Find Full Text PDF

Objective: Stereoelectroencephalography (SEEG) is increasingly being recognized as an important invasive modality for presurgical evaluation of epilepsy. This study focuses on the clinical and technical considerations of SEEG investigations when using conventional frame-based stereotaxy, drawing on institutional experience and a comprehensive review of relevant literature.

Methods: This retrospective observational study encompassed the surgical implantation of 201 SEEG electrodes in 16 epilepsy patients using a frame-based stereotactic instrument at a single tertiary-level center.

View Article and Find Full Text PDF

Late recognition of cardiac implantable electronic device misplacement in left ventricle: a case report.

Future Cardiol

April 2024

Department of Cardiology, School of Medicine, Imam Khomeini Hospital Complex, Tehran University of Medical Sciences, PO Box: 1419733141, Tehran, Iran.

Article Synopsis
  • * A chest x-ray later revealed the abnormal lead placement, and echocardiography confirmed it.
  • * The patient underwent surgical removal of the faulty lead, repair of the atrial septal defect, and correction of related issues, emphasizing the critical need for early diagnosis of complications to avoid severe outcomes.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!