Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Introduction: Optical coherence tomography (OCT) is a novel high-resolution imaging technique capable of visualizing in vivo structures at a resolution of ~10 μm. We have developed specialized OCT-based approaches that quantify diameter, speed, and flow rate in human cutaneous microvessels. In this study, we hypothesized that OCT-based microvascular assessments would possess comparable levels of reliability when compared with those derived using conventional laser Doppler flowmetry (LDF).
Methods: Speckle decorrelation images (OCT) and red blood cell flux (LDF) measures were collected from adjacent forearm skin locations on 2 d (48 h apart), at baseline, and after a 30-min rapid local heating protocol (30°C-44°C) in eight healthy young individuals. OCT postprocessing quantified cutaneous microvascular diameter, speed, flow rate, and density (vessel recruitment) within a region of interest, and data were compared between days.
Results: Forearm skin LDF (13 ± 4 to 182 ± 31 AU, P < 0.05) and OCT-derived diameter (41.8 ± 6.6 vs 64.5 ± 6.9 μm), speed (68.4 ± 9.5 vs 89.0 ± 7.3 μm·s), flow rate (145.0 ± 60.6 vs 485 ± 132 pL·s), and density (9.9% ± 4.9% vs 45.4% ± 5.9%) increased in response to local heating. The average OCT-derived microvascular flow response (pL·s) to heating (234% increase) was lower (P < 0.05) than the LDF-derived change (AU) (1360% increase). Pearson correlation was significant for between-day local heating responses in terms of OCT flow (r = 0.93, P < 0.01), but not LDF (P = 0.49). Bland-Altman analysis revealed that between-day baseline OCT-derived flow rates were less variable than LDF-derived flux.
Conclusions: Our findings indicate that OCT, which directly visualizes human microvessels, not only allows microvascular quantification of diameter, speed, flow rate, and vessel recruitment but also provides outputs that are highly reproducible. OCT is a promising novel approach that enables a comprehensive assessment of cutaneous microvascular structure and function in humans.