Purpose: The aim of this study was to present a case of a Baerveldt glaucoma implant lumen being completely occluded with a 3-0 Supramid stent suture.

Patient And Methods: The patient underwent Baerveldt glaucoma implant surgery with placement of an intraluminal 3-0 Supramid stent suture that acts to restrict flow across the device and reduce the risk of postoperative hypotony. Following suturing of the implant to the sclera, the device was flow tested. No flow was observed through the device tube and a significant ballooning of the tube diameter occurred with increased pressure on the device. The device was explanted from the eye and replaced with a different implant without further postoperative complication. The explanted device was assessed using custom microfluidic equipment in an in vitro environment.

Results: This phenomenon occurred despite using several different batches of the 3-0 Supramid stent suture and the device had to be removed and replaced with another device without complication. In vitro microfluidic assessment of the device demonstrated no flow across the device tube despite over 150 mm Hg of pressure being exerted on the device.

Conclusions: We hypothesize that the blockage occurred at the junction between the device tube and plate and that the ballooning phenomenon observed was due to a defect in the tube wall. This case highlights the importance of flow testing all glaucoma drainage devices before insertion given the variation in manufacturing conditions to avoid the risk of intraoperative complications.

Download full-text PDF

Source
http://dx.doi.org/10.1097/IJG.0000000000001198DOI Listing

Publication Analysis

Top Keywords

3-0 supramid
16
baerveldt glaucoma
12
glaucoma implant
12
supramid stent
12
device tube
12
device
11
stent suture
8
flow device
8
implant
5
flow
5

Similar Publications

Persistence of two coronaviruses and efficacy of steam vapor disinfection on two types of carpet.

Virol J

September 2024

Department of Food, Nutrition, and Packaging Sciences, Clemson University, 228A Life Science Facility, Clemson, SC, 29634, USA.

Background: Coronaviruses, a group of highly transmissible and potentially pathogenic viruses, can be transmitted indirectly to humans via fomites. To date, no study has investigated their persistence on carpet fibers. Establishing persistence is essential before testing the efficacy of a disinfectant.

View Article and Find Full Text PDF

Carpet cleaning guidelines currently do not include the use of an antimicrobial, except after a bodily fluid event. To address this gap, we compared the efficacy of three antimicrobials-two hydrogen peroxide-based (HO) products (A and B) and one chlorine-based product (C)-and a steam treatment against two norovirus surrogates, specifically feline calicivirus (FCV) and Tulane virus (TuV). These tests were performed on nylon carpets with either water-permeable or waterproof backing types.

View Article and Find Full Text PDF
Article Synopsis
  • Intraluminal placement of a 3-0 Supramid stent is shown to be an effective long-term solution for reversing hypotony (low eye pressure) after glaucoma drainage device (GDD) surgery, potentially reducing the need for additional surgical dissection.
  • In a study involving 9 patients, all experienced resolution of hypotony, with a total success rate of 7 out of 9 eyes after an average follow-up of 33.3 months.
  • Some patients required further interventions post-stent placement, but the stent procedure still presents a beneficial option for managing hypotony effectively.
View Article and Find Full Text PDF

The capability to produce suture material using three-dimensional (3D) printing technology may have applications in remote health facilities where rapid restocking of supplies is not an option. This is a feasibility study evaluating the usability of 3D-printed sutures in the repair of a laceration wound when compared with standard suture material. The 3D-printed suture material was manufactured using a fused deposition modelling 3D printer and nylon 3D printing filament.

View Article and Find Full Text PDF

Aim: To report the outcome of Baerveldt glaucoma implant (BGI) with Supramid ripcord use in neovascular glaucoma (NVG).

Methods: We retrospectively evaluated the surgical outcome of the BGI with Supramid 3/0 ripcord stent in patients with NVG. No tube ligation or venting slits were performed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!