Background: Radiation risk minimisation during cardiac catheterisation, particularly atrial fibrillation (AF) ablation procedures, requires a multifaceted approach involving both technique and technology.

Objective: To evaluate radiation dose associated with cryoballoon compared to conventional radiofrequency (RF) AF ablation procedures.

Methods: Consecutive patients undergoing AF ablation were collected in a single-centre registry. Radiation and procedural data for index AF ablation procedures using four ablation techniques were compared: RF ablation with and without the anti-scatter grid (RF Grid and RF Gridless), and cryoballoon ablation utilising first and second-generation Medtronic Arctic Front Cryoballoons (Cryo; Arctic Front, Medtronic, Minneapolis, MN, USA), with and without 3D imaging during fluoroscopy (Cryo 3D and Cryo).

Results: We studied 418 patients; 30 RF Grid, 68 RF No Grid, 12 Cryo 3D and 308 Cryo. The dose area product (DAP; Gycm), adjusted for patient standardised weight of 80kg, was significantly higher for the Cryo 3D (21.91) and RF Grid (7.31) than the Cryo (2.13) and RF Gridless (3.31) groups; as a result, Cryo 3D was discontinued and RF Grid was only used when clinically required. Comparing the remaining groups, DAP for Cryo was significantly lower than RF Gridless (p<0.001) mainly attributable to a difference in fluoroscopy use (13.2 vs. 17.3 mins; p<0.001). Cryo procedure time was also significantly shorter (80 mins vs. 133 mins; p<0.01).

Conclusion: Cryoballoon AF ablations can be performed efficiently using Gridless fluoroscopy techniques achieving significant reduction in radiation exposure and procedure time while maintaining safety and efficacy. In this observational study, cryoballoon AF ablations compare favourably to conventional RF ablations in terms of these parameters.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hlc.2018.11.015DOI Listing

Publication Analysis

Top Keywords

ablation
8
cryoballoon ablation
8
ablation procedures
8
grid grid
8
arctic front
8
cryo
8
grid
6
radiation
4
radiation exposure
4
exposure optimisation
4

Similar Publications

 Combination therapy, which synergistically enhances treatment efficacy and inhibits disease progression through the combined effects of multiple drugs, has emerged as a mainstream approach for treating complex diseases and alleviating symptoms. However, drug-drug interactions (DDIs) can sometimes lead to adverse reactions, potentially endangering lives. Therefore, developing efficient and accurate DDI prediction methods is crucial for elucidating drug mechanisms and preventing side effects.

View Article and Find Full Text PDF

Pudendal Neuralgia: A Review of the Current Literature.

Curr Pain Headache Rep

January 2025

Department of Anesthesia, Division of Pain Medicine, University of Virginia, Charlottesville, VA, USA.

Purpose Of Review: This paper aims to review pudendal neuralgia pathophysiology, risk factors, diagnosis, and treatment options.

Recent Findings: Conservative and pharmacologic options are first line treatments for the treatment of pudendal neuralgia. Interventional treatment such as, pudendal nerve blocks can be tried if first line treatments feel to provide adequate analgesia.

View Article and Find Full Text PDF

Bipolar ablation's safety and efficacy: reassurance coming from ex vivo.

J Interv Card Electrophysiol

January 2025

Division of Cardiovascular Medicine, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, GA, USA.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!