"Peroperative estimation of bone quality and primary dental implant stability".

J Mech Behav Biomed Mater

Institute for Surgical Technology and Biomechanics, University of Bern, Stauffacherstrasse 78, 3014 Bern, Switzerland; Institute for Mechanical, Process and Energy Engineering, Heriot-Watt University, UK EH14 4AS, Edinburgh, United Kingdom.

Published: April 2019

Objectives: Dental implants are widely used to restore function and appearance. It may be essential to choose the appropriate drilling protocol and implant design in order to optimise primary stability. This could be achieved based on an assessment of the implantation site with respect to bone quality and objective biomechanical descriptors such as stiffness and strength of the bone-implant system. The aim of this ex vivo study is to relate these descriptors with bone quality, with a pre-implantation indicator of implant stability: pilot-hole drilling force (F), and with two post-implantation indicators: maximal implantation torque (T) and resonance frequency analysis (RFA).

Methods: Eighty trabecular bone specimens were cored from human vertebrae and bovine tibiae. Bone volume fraction (BV/TV), a representative for bone quality, was obtained through micro-computed tomography scans. Implants were kept in controlled laboratory conditions following standard surgical procedures. Forces and torques were recorded and RFA was assessed after implantation. Off-axis compression tests were conducted on the implants until failure. Implant stability was identified by stiffness and ultimate force (F). The relationships between BV/TV, Stiffness, F and F, T, RFA were established.

Results: F correlated well with BV/TV of the implantation site (r = 0.81), stiffness (r = 0.75) and F (r = 0.80). T correlated better with stiffness (r = 0.86) and F (r = 0.94) than RFA (r = 0.77 and r = 0.74, respectively).

Conclusion: Our results indicate that BV/TV and bone-implant stability can be directly estimated by the force needed for the pilot drilling that occurs during the site preparation before implantation. Moreover, implantation torque outperforms RFA for evaluating the mechanical competence of the bone-implant system.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jmbbm.2018.12.035DOI Listing

Publication Analysis

Top Keywords

bone quality
16
implantation site
8
bone-implant system
8
implant stability
8
implantation torque
8
bone
6
implantation
6
stiffness
5
"peroperative estimation
4
estimation bone
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!