Background: Extracorporeal shock wave lithotripsy (ESWL) is considered as the method of choice for treatment of most stone diseases. The present study aims to evaluate radiation dose received by patients undergoing ESWL.

Methods: In total, 46 patients from both genders were referred to the Shohadaye Ashayer hospital of Khorramabad, Iran and were included in the present study. Patients were positioned in anteroposterior (AP) projection and along 30° anterior oblique (AO) projection upon the X-ray fluoroscopy table and exposures were conducted. Thermoluminescent dosimeters (TLDs) were used for radiation dose measurements. To evaluate the entrance surface dose (ESD), each TLD chip was taped on back of patient at the entrance surfaces of the X-ray beam.

Results: The mean number of stones in each patient was 1.4. The stone sizes ranged from 7 to 29 mm, and a mean of 3200 pulses were need for each patient. The mean ESD in the postero-anterior (PA) and oblique X-ray beam entrance were obtained at 12.04 and 68.84 mGy, respectively. There was linear correlation between patient dose with fluoroscopy time, tube current (mA), tube potential (kVp) and patient position (P < 0.001); however, we found no strong correlation between patient dose with patient body mass (P = 0.837), number of shock wave pulses (P = 0.089), stone size (P = 0.773) and locations (P = 0.463).

Conclusion: The data obtained in the current study are comparable with information available in the literature. They emphasized that ESWL exposes patients to much more radiation compared to those from conventional radiography and is in the range of computed tomography (CT) procedures. Therefore, following safety guidelines is recommended.

Download full-text PDF

Source

Publication Analysis

Top Keywords

radiation dose
12
shock wave
12
dose received
8
received patients
8
extracorporeal shock
8
wave lithotripsy
8
correlation patient
8
patient dose
8
patient
7
dose
6

Similar Publications

Purpose: Immune checkpoint blockades (ICBs) are promising, however they do not fit all types of tumor, such as those lack of tumor antigens. Induction of potent anti-tumor T cell immunity is critical for cancer therapy. In this study, we investigated the efficacy of immunotherapy via the immunogenic cell death (ICD) dying tumor cells in mouse models of lung metastasis and tumorigenesis.

View Article and Find Full Text PDF

A systematic review of the effectiveness of leaded glasses for ensuring safety among healthcare professionals in fluoroscopy.

J Med Imaging Radiat Sci

January 2025

Instituto Politécnico de Coimbra, ESTESC - Coimbra Health School, Medical Imaging and Radiotherapy, Rua 5 de Outubro, S. Martinho do Bispo, Coimbra 3046-854, Portugal. Electronic address:

Background: Currently, there is an increase in procedures across various clinical specialties involving the use of ionising radiation.

Objective: The primary objective of this systematic review is to analyse and compare the existing literature regarding the effectiveness of leaded glasses for healthcare professionals.

Methods: Comprehensive literature searches were conducted for relevant studies published between 2018 and 2023 using the Scopus, PubMed, and Web of Science databases according to preferred reporting items for systematic reviews and meta-analyses (PRISMA) methodology.

View Article and Find Full Text PDF

Purpose: To train and validate KB prediction models by merging a large multi-institutional cohort of whole breast irradiation (WBI) plans using tangential fields.

Methods: Ten institutions (INST1-INST10, 1481 patients) developed their KB-institutional models for left/right WBI (ten models for right and eight models for left). The transferability of models among centers was assessed based on the overlap of the geometric Principal Component (PC1) of each model when applied to other institutions and/or on the presence of significantly different optimization policies.

View Article and Find Full Text PDF

Multi-layer shielding optimization of a high activity Am-Be mixed field irradiation facility.

Appl Radiat Isot

January 2025

Experimental Nuclear Physics Department, Nuclear Research Centre, Egyptian Atomic Energy Authority, Egypt; Cyclotron Facility, Egyptian Atomic Energy Authority, Egypt.

Neutron and gamma-ray shielding design for a 30Ci (1.11TBq) Am-Be irradiation facility is studied using MCNP5 Monte Carlo simulation code. The study focuses on the optimization of the shielding layers of the previously planned neutron irradiation facility.

View Article and Find Full Text PDF

Objective: The study aims to systematically characterize the effect of CT parameter variations on images and lung radiomic and deep features, and to evaluate the ability of different image harmonization methods to mitigate the observed variations.

Approach: A retrospective in-house sinogram dataset of 100 low-dose chest CT scans was reconstructed by varying radiation dose (100%, 25%, 10%) and reconstruction kernels (smooth, medium, sharp). A set of image processing, convolutional neural network (CNNs), and generative adversarial network-based (GANs) methods were trained to harmonize all image conditions to a reference condition (100% dose, medium kernel).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!