Objective: The purpose of this study was to compare image quality and lesion detection capability between a digital and an analog PET/CT system in oncological patients.

Materials And Methods: One hundred oncological patients (62 men, 38 women; mean age of 65 ± 12 years) were prospectively included from January-June 2018. All patients, who accepted to be scanned by two systems, consecutively underwent a single day, dual imaging protocol (digital and analog PET/CT). Three nuclear medicine physicians evaluated image quality using a 4-point scale (-1, poor; 0, fair; 1, good; 2, excellent) and detection capability by counting the number of lesions with increased radiotracer uptake. Differences were considered significant for a p value <0.05.

Results: Improved image quality in the digital over the analog system was observed in 54% of the patients (p = 0.05, 95% CI, 44.2-63.5). The percentage of interrater concordance in lesion detection capability between the digital and analog systems was 97%, with an interrater measure agreement of κ = 0.901 (p < 0.0001). Although there was no significant difference in the total number of lesions detected by the two systems (digital: 5.03 ± 10.6 vs. analog: 4.53 ± 10.29; p = 0.7), the digital system detected more lesions in 22 of 83 of PET+ patients (26.5%) (p = 0.05, 95% CI, 17.9-36.7). In these 22 patients, all lesions detected by the digital PET/CT (and not by the analog PET/CT) were < 10 mm.

Conclusion: Digital PET/CT offers improved image quality and lesion detection capability over the analog PET/CT in oncological patients, and even better for sub-centimeter lesions.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00259-019-4260-zDOI Listing

Publication Analysis

Top Keywords

image quality
12
digital analog
12
analog pet/ct
12
quality lesion
8
lesion detection
8
detection capability
8
comparison image
4
detection digital
4
pet/ct objective
4
objective purpose
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!