Background and Purpose- The safety and efficacy of mechanical thrombectomy in patients with acute ischemic stroke has been demonstrated. However, the impact of stent retriever size on clinical and angiographic outcomes is not well established. Methods- This was a retrospective ad hoc analysis of data from the STRATIS (Systematic Evaluation of Patients Treated With Neurothrombectomy Devices for Acute Ischemic Stroke) registry-a prospective, multicenter study of patients with large vessel occlusion treated with the Solitaire stent retriever. An independent core laboratory, blinded to clinical outcomes, reviewed all procedures and angiographic data to classify stent retriever size, targeted clot location, recanalization after each pass, and the number of stent retriever passes. The primary angiographic end point was first-pass effect (FPE) as determined by a core laboratory and defined as achieving near-complete revascularization (modified Thrombolysis in Cerebral Infarction ≥2c) after the first pass without the use of rescue therapy. Rates of modified FPE were also assessed, defined as meeting all criteria for FPE but achieving modified Thrombolysis in Cerebral Infarction ≥2b after first pass. The primary clinical end point was functional independence (modified Rankin Scale, 0-2) at 3 months as determined on-site. Outcome comparisons were made across the stent retriever size groups and adjusted for baseline characteristics. Results- Of 715 patients, a 4×20 stent retriever was used in 201 (28%) patients, 4×40 was used in 270 (38%) patients, and 6×30 was used in 244 (34%) patients. The 4×40 group had the highest rate of FPE ( P=0.003 versus 6×30) and modified FPE ( P=0.038 versus 4×20; P=0.0001 versus 6×30). Final revascularization was not significantly different across the groups, and there were no significant differences in functional dependence or mortality at 90 days post-procedure. Use of the longer stent retriever (4×40) was an independent predictor of achieving modified FPE ( P=0.037 versus 6×30; P=0.037 versus 4×20). Conclusions- The longer stent retriever (4×40) demonstrated the highest rate FPE and modified FPE compared with larger diameter or shorter stent retrievers, suggesting that their routine use may improve early revascularization success. Clinical Trial Registration- URL: https://www.clinicaltrials.gov . Unique identifier: NCT02239640.

Download full-text PDF

Source
http://dx.doi.org/10.1161/STROKEAHA.118.022987DOI Listing

Publication Analysis

Top Keywords

stent retriever
36
retriever size
16
modified fpe
16
versus 6×30
12
retriever
9
stent
9
impact stent
8
size clinical
8
clinical angiographic
8
angiographic outcomes
8

Similar Publications

Background And Purpose: The safety and effectiveness of endovascular techniques in elderly patients with large vessel occlusion (LVO) remain controversial. We investigated the angiographic and clinical outcomes of nonagenarians treated with different endovascular techniques using a balloon guide catheter (BGC), distal aspiration catheter (DAC), and/or stent retriever (SR).

Methods: We analyzed the data from the Registry of Combined versus Single Thrombectomy Techniques (ROSSETTI) of consecutive nonagenarian patients with anterior circulation LVO and compared the outcomes of those treated with BGC+noDAC+SR (101-group), BGC+DAC+SR (111-group), and noBGC+DAC+SR (011-group).

View Article and Find Full Text PDF

Background And Purpose: Endovascular thrombectomy (EVT) for acute ischemic stroke (AIS) with M2 segment occlusion of the middle cerebral artery (MCA) is debatable. This study assessed the efficacy, safety, and functional outcomes of EVT in M2 occlusion patients, examining differences in outcomes based on the dominance of the occluded segment (DomM2 vs. Non-DomM2).

View Article and Find Full Text PDF

Impact of thrombus composition on virtual thrombectomy procedures using human clot analogues mechanical data.

J Mech Behav Biomed Mater

January 2025

Department of Chemistry, Materials and Chemical Engineering 'Giulio Natta', Politecnico di Milano, Italy.

Endovascular thrombectomy (EVT) aims at restoring blood flow in case of acute ischemic stroke by removing the thrombus occluding a large cerebral artery. During the procedure with stent-retriever, the thrombus is captured within the device, which is then retrieved, subjecting the thrombus to several forces, potentially leading to its fragmentation. In silico studies, along with mechanical characterisation of thrombi, can enhance our understanding of the EVT, helping the development of new devices and interventional strategies.

View Article and Find Full Text PDF

Background: There are limited therapeutic options in cases of failed reperfusion (modified thrombolysis in cerebral infarction [mTICI] score < 2b) after stent-retriever and/or aspiration based endovascular treatment (EVT) for acute ischemic stroke. Despite the absence of data supporting its use, rescue therapy (balloon angioplasty and/or stent implantation) is often utilized in such cases. Studies are limited to large vessel occlusions, while the outcomes and complications after rescue therapy in medium/distal vessel occlusions (MDVOs) have not been reported.

View Article and Find Full Text PDF

Outcomes and Complications Associated with Mechanical Thrombectomy in the Treatment of Acute Ischemic Stroke.

Cardiol Cardiovasc Med

December 2024

Department of Translational Research, College of Osteopathic Medicine of the Pacific, Western University of Health Sciences, Pomona CA 91766, USA.

Universally, stroke presents as neurological deficits due to the obstruction of blood supply to specific regions of the brain. Among the three main categories of stroke, acute ischemic stroke is the leading cause of death and disability worldwide. As of today, there are two effective treatment methods: thrombolysis and endovascular therapy.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?