Background: Endotracheal suctioning is mandatory to prevent complications caused by the retention of tracheal secretions. Endotracheal suctioning is often performed late, when patients show signs of respiratory and hemodynamic alterations. We conceived a prototype device that, when synchronized with the ventilator, automatically removes secretions collected below the endotracheal tube (ETT) cuff, thus avoiding endotracheal suctioning. The aim of our investigation was to assess the performance of this novel prototype in vitro.

Methods: Three studies were performed to examine the characteristics of the prototype. We tested device's ability to generate an effective artificial cough flow (artificial cough maneuver) > 1 L/s by rapidly deflating the ETT cuff within the time of a sustained inflation (at 30 and at 40 cm HO) (cough flow study). We also tested the prototype's ability to remove the fluid positioned below the ETT cuff using saline dye (fluid removal study), and to prevent the aspiration of saline dye from above the ETT cuff during the deflation phase of the ETT cuff (aspiration study). The trachea model was positioned at 45° in the aspiration study, and horizontally in the other two studies.

Results: In the cough flow study, the prototype provided an effective artificial cough maneuver, with a mean ± SD of 1.78 ± 0.19 L/s (range, 1.42-2.14 L/s). The tracheal pressure after ETT cuff deflation never decreased below the PEEP level. In the fluid removal study, the prototype cleared the fluid from below the ETT cuff and the experimental trachea. No fluid was aspirated from the area above the ETT cuff toward the lower airways.

Conclusions: We conceived an system capable of automatically expelling fluid from below the ETT cuff outside an experimental trachea by generating an artificial cough maneuver. This system may decrease the use of endotracheal suctioning and its complications. Future in vivo studies are needed to confirm this first in vitro evaluation.

Download full-text PDF

Source
http://dx.doi.org/10.4187/respcare.06076DOI Listing

Publication Analysis

Top Keywords

ett cuff
36
artificial cough
20
cough maneuver
16
endotracheal suctioning
16
cough flow
12
cuff
10
ett
9
secretions endotracheal
8
endotracheal tube
8
effective artificial
8

Similar Publications

Background: Monitoring core temperature is important for patients under anesthesia. Esophageal and pulmonary artery blood temperatures can be used for measuring core temperature during general anesthesia. However, these methods pose challenges, especially when the placement of an esophageal thermometer and pulmonary artery catheter (PAC) is either impractical or not the preferred approach.

View Article and Find Full Text PDF

A 'can't intubate, can't oxygenate' (CICO) situation is an uncommon and time-critical emergency. Many institutions have adopted a 'scalpel-bougie-endotracheal tube (ETT)' technique based on evidence produced by the 4th National Audit Project of the Royal College of Anaesthetists and 2015 Difficult Airway Society guidelines. We made a modification to the traditional 'scalpel-bougie-ETT' technique, using a shortened bougie and replacing the ETT with a cuffed Melker airway in a preassembled device (called 'Secure Airway for Front-of-neck Emergencies' (SAFE airway device)), which we felt might reduce cognitive load on a single operator in an emergency CICO situation.

View Article and Find Full Text PDF

Background: The optimal endotracheal tube (ETT) cuff pressure remains contentious. In the traditional consideration that the level 30 cmHO is considered safe, balancing the prevention of reflux aspiration against airway mucosal damage. Whether this pressure level can cause potential damage to the airway mucosa remains to be discussed.

View Article and Find Full Text PDF

Design considerations for development of cuffed endotracheal tube for small airways.

Paediatr Anaesth

February 2025

Research and Development, Acute Care & Monitoring, Medtronic, Athlone, Ireland.

Article Synopsis
  • * Neonates and pediatric patients have specific airway structures and functional needs that influence the design of ETTs, making it important to cater to these requirements.
  • * While historically uncuffed ETTs were preferred for children under eight, advancements in cuffed ETT design now allow for safer use in smaller, more delicate airways.
View Article and Find Full Text PDF

Background: Mechanical ventilation is a common and often lifesaving intervention that is utilised in intensive care. However, the practices can vary between centres. Through this national survey we aim to gain more information about different strategies adopted across the UK.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!