Background: There is a lack of evidence in reliability of the modified sphygmomanometer to measure the brace-interface pad pressure in adolescents with Scheuermann's kyphosis (SK).

Objectives: To evaluate the intra- and inter-observer reliability of modified sphygmomanometer for measuring interface pad pressure in the Milwaukee brace.

Methods: Two observers measured the pressure of the shoulder and kyphosis pads in Milwaukee brace on 33 adolescents with SK with an average age of 14.67 ± 1.72 years and Cobb angle 64.48∘± 7.53∘. All measurements were obtained in one day. The measurements were done in the inhalation and exhalation of tidal breathing during standing and sitting positions. The intraclass correlation coefficient (ICC), 95% confidence interval (CI), and linear mixed model ANOVA effects were calculated.

Results: The intra-observer reliability varied from a good ICC of 0.81 (0.66-0.90) to an excellent of 0.97 (0.95-0.98). The inter-observer reliability also varied from a good ICC of 0.82 (0.67-0.91) to an excellent of 0.96 (0.93-0.98). The linear mixed model ANOVA analysis showed that the rater, position, and breathing had a significant effect on the pad pressure measurements.

Conclusions: The modified sphygmomanometer is a reliable tool to measure the pad pressure of the Milwaukee brace for adolescents with SK.

Download full-text PDF

Source
http://dx.doi.org/10.3233/BMR-171041DOI Listing

Publication Analysis

Top Keywords

pad pressure
16
inter-observer reliability
12
milwaukee brace
12
brace adolescents
12
modified sphygmomanometer
12
intra- inter-observer
8
reliability modified
8
pressure milwaukee
8
linear mixed
8
mixed model
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!