Objective: This study reports the technical aspects and 30-day outcomes of the prospective, multicenter early feasibility study designed to evaluate the GORE EXCLUDER Thoracoabdominal Branch Endoprosthesis (TAMBE; W. L. Gore & Associates, Flagstaff, Ariz).
Methods: Thirteen patients with pararenal or extent IV thoracoabdominal aortic aneurysms were prospectively enrolled at five U.S sites and one non-U.S. site from 2014 to 2016. The TAMBE included four portals with either retrograde or antegrade renal portal configuration and used GORE VIABAHN Balloon-Expandable Endoprosthesis (W. L. Gore & Associates) for stenting of the renal and mesenteric arteries. The primary end point was procedural safety at 30 days, defined by absence of major adverse events, including any-cause mortality, myocardial infarction, stroke, paraplegia, bowel ischemia, respiratory failure, severe acute kidney injury (>50% decline in estimated glomerular filtration rate), dialysis, and procedural blood loss >1000 mL.
Results: There were 11 male and two female patients with a mean age of 69 ± 8 years. Mean aneurysm diameter was 61 ± 13 mm. A total of 52 renal and mesenteric arteries were incorporated (4 vessels/patient). Technical success was achieved in 12 patients (92%). One patient had inadvertent occlusion of a right renal artery due to dissection. There was no mortality, aneurysm rupture, conversion to open repair, dialysis, or spinal cord injury. Mean length of hospital stay was 5 ± 3 days. At 30 days, four patients (31%) had major adverse events, all due to procedural blood loss >1000 mL. One patient had a type I endoleak at the distal renal branch, which was successfully treated by placement of an additional renal stent before dismissal. Computed tomography angiography at 30 days showed patent target vessels and no type I or type III endoleak.
Conclusions: This study confirms the early feasibility of the TAMBE for treatment of pararenal and extent IV thoracoabdominal aortic aneurysms. The high technical success, no mortality, and low morbidity rate support continuation of clinical investigation in a larger population of patients.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jvs.2018.10.103 | DOI Listing |
PLoS One
January 2025
School of Public Health, Makerere University, Kampala, Uganda.
Introduction: Cervical cancer (CC) is the leading cause of cancer-related deaths among Uganda women, yet rates of CC screening are very low. Training women who have recently screened to engage in advocacy for screening among women in their social network is a network-based strategy for promoting information dissemination and CC screening uptake.
Methods: Drawing on the Exploration, Preparation, Implementation and Sustainment (EPIS) framework for implementation science, this hybrid type 1 randomized controlled trial (RCT) of a peer-led, group advocacy training intervention, Game Changers for Cervical Cancer Prevention (GC-CCP), will examine efficacy for increasing CC screening uptake as well as how it can be implemented and sustained in diverse clinic settings.
PLoS One
January 2025
Department of Biomedical and Robotics Engineering, Incheon National University, Incheon, Korea.
Gait disturbance is one of the most common symptoms in patients with Parkinson's disease (PD) that is closely associated with poor clinical outcomes. Recently, video-based human pose estimation (HPE) technology has attracted attention as a cheaper and simpler method for performing gait analysis than marker-based 3D motion capture systems. However, it remains unclear whether video-based HPE is a feasible method for measuring temporospatial and kinematic gait parameters in patients with PD and how this function varies with camera position.
View Article and Find Full Text PDFJ Endourol
January 2025
Department of Urology, Peking University First Hospital, Institution of Urology, Beijing Key Laboratory of Urogenital Diseases (Male) Molecular Diagnosis and Treatment Center, National Urological Cancer Center, Peking University, Beijing, China.
The KangDuo Surgical Robot-1500 (KD-SR-1500) is a newly developed surgical robot. We aim to evaluate the feasibility and efficiency of the KD-SR-1500 system for robot-assisted radical prostatectomy (RARP). This prospective, multicenter, single-arm clinical study was conducted among 18-75-year-old patients with suspected T1-2N0M0 prostate cancer scheduled for RARP.
View Article and Find Full Text PDFJ Funct Biomater
January 2025
Department of Mechanics, Materials and Biomedical Engineering, Faculty of Mechanical Engineering, Wrocław University of Science and Technology, Smoluchowskiego 25, 50-372 Wrocław, Poland.
This study focuses on the development and evaluation of the OrthoNail hybrid intramedullary implant for lower limb lengthening in patients requiring significant skeletal reconstruction. The implant addresses the challenges in load-bearing during rehabilitation, providing a robust solution that is capable of supporting physiological loads. Mechanical tests, including axial compression, tension, torsion, and 3,4-point bending, determined the implant's load capacity and fatigue resistance, while finite element analysis assessed stress distributions in bone tissue and around screw holes during single-leg stance, with boundary conditions derived from Orthoload database data.
View Article and Find Full Text PDFCurr Oncol
December 2024
Department of Radiation Oncology, China Medical University Hospital, Taichung City 404, Taiwan.
Background: Since 2023, ChatGPT-4 has been impactful across several sectors including healthcare, where it aids in medical information analysis and education. Electronic patient-reported outcomes (ePROs) play a crucial role in monitoring cancer patients' post-treatment symptoms, enabling early interventions. However, managing the voluminous ePRO data presents significant challenges.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!