Objective: To assess the radiation dose and image quality of ultra-low dose (ULD)-CT colonography (CTC) obtained with the combined use of automatic tube current (mAs) modulation with a quality reference mAs of 25 and sinogram-affirmed iterative reconstruction (SAFIRE), compared to low-dose (LD) CTC acquired with a quality reference mAs of 55 and reconstructed with filtered back projection (FBP).

Methods: Eighty-two patients underwent ULD-CTC acquisition in prone position and LD-CTC acquisition in supine position. Both ULD-CTC and LD-CTC protocols were compared in terms of radiation dose [weighted volume computed tomography dose index (CTDI ) and effective dose], image noise, image quality, and polyp detection.

Results: The mean effective dose of ULD-CTC was significantly lower than that of LD-CTC (0.98 and 2.69 mSv respectively, P < 0.0001) with an overall dose reduction of 63.2%. Image noise was comparable between ULD-CTC and LD-CTC (28.6 and 29.8 respectively, P = 0.09). There was no relevant difference when comparing image quality scores and polyp detection for both 2D and 3D images.

Conclusion: ULD-CTC allows to significantly reduce the radiation dose without meaningful image quality degradation compared to LD-CTC.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6333183PMC
http://dx.doi.org/10.1002/acm2.12510DOI Listing

Publication Analysis

Top Keywords

image quality
12
ultra-low dose
8
automatic tube
8
tube current
8
sinogram-affirmed iterative
8
iterative reconstruction
8
radiation dose
8
quality reference
8
reference mas
8
quality
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!